background image

®

Wireless Load Cell ±50N

PS

-

3216

10

013-15734A

Appendix B: Calibration

Calibration is not necessary, especially if you are 

measuring a change in force rather than absolute force 

values. However, it is possible to calibrate the sensor. 

Prepare for Calibration

Calibration will need a 1 kilogram mass, the hook 

attachment, and a horizontally mounted support rod to 

hold the sensor. The sensor will need to be “connected” 

with a tablet or computer, and the data collection 

software (for example, SPARKvue) should be running.

Using SPARKvue Software for Calibration

See the SPARKvue Help for information about 

calibrating: 

In SPARKvue, select the HELP button (

) in any 

screen, such as the Home Screen.

1. Mount the Wireless Load Cell on a calibration fix

-

ture. 

2. Click (or press) the Experiment Tools button ( ).

The Experiment Tools screen opens.

3. Click Calibrate Sensor.

The Calibrate Sensor: Select Measurement screen 

opens.

4. Click the Sensor box and click the sensor to be 

calibrated.

5. Click the Calibration Type box and click a calibration 

type. (For this example, click “2-point”.)

6. Click Next.

The Calibrate Sensor Enter Values screen opens.

7. Hang the 1 kilogram mass on the calibration fixture. 

8. The gravitational force on the mass is pulling in the 

negative direction at -9.8 newtons (N). Under 

Calibration Point 1, click the Standard Value box 

and enter the known force value (that is, -9.8).

9. Under Calibration Point 1, click Read From Sensor.

The value measured by the sensor is transferred to 

the Sensor Value box.

10. Remove the 1 kilogram. The net force now is zero N. 

11. Under Calibration Point 2, click the Standard Value 

box and enter the second known force value (i.e., 0).

12. Under Calibration Point 2, click the Read From 

Sensor box.

The second value measured by the sensor is 

transferred to the Sensor Value box.

13. Click OK.

Using PASCO Capstone for Calibration

See the PASCO Capstone Help for information about 

calibrating:

In PASCO Capstone, select PASCO Capstone Help 

from the Help menu, or press F1.

1. Mount the Wireless Load Cell on a calibration fix

-

ture.

2. Click Calibration ( ) in the Tools palette.

3. Choose the measurement you would like to calibrate 

now: Force Measurement.

4. Click Next.

5. Choose the type of calibration you would like to 

perform: Two Standards (2 point).

6. Click Next.

Summary of Contents for PS-3216

Page 1: ...The PASCO Data Collection Software can show the compression or tension forces the accelera tion in X Y and Z and the resultant acceleration The sensor uses a micro electromechanical system MEMS device...

Page 2: ...he sensor off press and hold the ON OFF button for a moment until the Battery Status LED shines red The Wireless Load Cell puts itself to sleep after several minutes of inactivity if not connected and...

Page 3: ...tton All of the Force and Acceleration measurements will be set to zero To Zero Force or Acceleration Another option is to set either the Force measurement or the Acceleration measurements to zero Sta...

Page 4: ...option is to use the Common Rate button Click the Common Rate button in the Control palette below the workbook to open the Common Rate menu Select Wireless Force Sensor from the menu and then click th...

Page 5: ...th the PASCO Structures Systems such as the ME 3581 Building Better Bridges Kit The sensor can measure compression and tension forces in any member of the PASCO Structures Systems Use thumbscrews to m...

Page 6: ...Y Z Orientation In this orientation of the sensor the X axis direction is horizontal the Y axis direction is vertical and and the Z axis direction is perpendicular to the label on the front of the sen...

Page 7: ...rtant to making the sensor simple and always ready to use so all of the PASCO wireless products are designed for long battery life For example the sensor turns itself off after a few minutes of inacti...

Page 8: ...ence and 2 this device must accept any interference received including interference that may cause undesired operation CE Statement This device has been tested and found to comply with the essential r...

Page 9: ...hing lower than 0x6 means an older version of Bluetooth Your device will need the PS 3500 USB Bluetooth 4 0 Adapter 1The Mac Mini and MacBook Air were updated with Bluetooth SMART support in 2011 The...

Page 10: ...this example click 2 point 6 Click Next The Calibrate Sensor Enter Values screen opens 7 Hang the 1 kilogram mass on the calibration fixture 8 The gravitational force on the mass is pulling in the neg...

Page 11: ...linear equation that transforms the raw input data into the calibrated output data The linear function is of the form Raw Input Slope x Calibrated Output Offset Or Calibrated Output Raw Input Offset...

Page 12: ...e shifts so that it intersects the new point but its slope does not change Offset calibration is usually used to make one sensor agree with another sensor Due to normal variation among probes a second...

Reviews: