background image

C h e m i s t r y   M u l t i M e a s u r e   S e n s o r

Sp e c if ic a t i o n s

8

012-09523B

Appendix A: PASCO Capstone Calibration

In PASCO Capstone, click the Hardware Setup icon in the tools palette to open the Hardware Setup panel. Confirm that the 
icon of the sensor appears with the icon of the interface being used. Click the Hardware Setup icon again to close the panel. 
Click the Calibration icon (

) in the Tools palette to open the Calibration panel. For a two-point calibration, you will need 

two “known standards,” such as two buffer solutions of known pH or two liquids of known temperature, such as ice water and 
hot water. For temperature calibration, a standard thermometer will also be needed.

1.

The first step in the Calibration panel is to choose the type of mea-
surement you would like to calibrate. The default for the Chemistry 
sensor is “Temperature”. The other choice on the menu is “pH”. 
Click “Next”. 

2.

In the second step, the temperature probe is automatically selected (if 
the default measurement is kept). Move to the third step.

3.

The third step in the panel is to choose the type of calibration. “Two 
Standards (2 point)” is the default. Click “Next”. 

4.

The fourth step in the panel is to calibrate the first point. The “Stan-
dard Value” shows 0.00 °C. (This Standard Value can be changed if 
necessary. Highlight the “Standard Value” and enter a new value if 
needed). For this example, put the temperature probe into a “known 
standard” of ice water. Click “Set Current Value to Standard Value” 
and then click “Next”. 

5.

The fifth step in the panel is to calibrate the second point. The “Stan-
dard Value” shows 100 °C. For this example, put the temperature 
probe into a “known standard” of hot water along with the thermom-
eter. Highlight the “Standard Value” and enter the temperature from 
the thermometer. Click “Set Current Value to Standard Value” and 
then click “Next”.

6.

The sixth step in the panel is to review and accept the calibration 
information. Click “Back” to return to a previous step. Otherwise, 
click “Finish”. The first step in the panel appears again. Follow the 
same procedure to calibrate for another type of measurement.

7.

Click the Calibration icon in the Tools palette to close the Calibration 
panel. 

Summary of Contents for PASPort PS-2170

Page 1: ...yurethane Tubing 60 cm 0 125 inch ID 6 Quick release Connectors qty 4 Hose and Coupling Kit 7 Tubing Connectors qty 4 8 Syringe 60 cc 9 Alligator Clip with Insulator qty 2 each Polyurethane Tubing blu...

Page 2: ...RP port BNC port Align the grooves of the BNC connector on the probe with the pins on the BNC port Push the connector onto the port and turn the connector clockwise until it locks in place If you plan...

Page 3: ...easurement for that display Using SPARKvue HD Connect a PASPORT compatible interface to the computer and start the SPARKvue HD software Refer to the documentation that came with your PASPORT interface...

Page 4: ...measurement can be displayed in units of C F or K You can use the included stainless steel probe or a different probe such as the Fast response Temperature Probe PS 2135 or Skin Surface Temperature Pr...

Page 5: ...Electrode Maintenance and Storage Cleaning If the pH electrode becomes contaminated use one of these methods to clean and restore it After any of these procedures rinse the electrode with deionized w...

Page 6: ...tors to connect the tubing to larger diameter tubing or to a rubber stopper with a hole Note Do not allow liquid to enter the pressure port Voltage Use the voltage component of the multi sensor to mea...

Page 7: ...Model No PS 2170 Specifications 7 012 09523B Voltage Range 10 V to 10 V Accuracy 0 1 V Resolution 0 04 mV Voltage Protection up to 240 V Input Resistance 2 M...

Page 8: ...ept Move to the third step 3 The third step in the panel is to choose the type of calibration Two Standards 2 point is the default Click Next 4 The fourth step in the panel is to calibrate the first p...

Page 9: ...he Graph display screen to open the Experiment Tools menu In the Experiment Tools menu touch Calibrate Sensor The screen will show Calibrate Sensor Select Measure ment The Sensor Chemistry Measurement...

Page 10: ...be in distilled water 5 For Calibration Point 2 touch the down arrow to scroll down to the bottom of the screen Place the probe s sensor into the second known standard such as the pH 7 buffer solution...

Page 11: ...s pH Probe into the first known standard pH 4 buf fer solution Wait a few moments and then press F3 H to Read Pt 1 5 Use the down arrow to scroll down to the second calibration point Highlight Pt 2 mV...

Page 12: ...products or service names are or may be trademarks or service marks of and are used to identify products or services of their respective owners For more information visit www pasco com legal Teflon i...

Reviews: