background image

M o d e l   N o .   P S - 2 1 7 0

 S p e c i f i c a t i o n s

9

012-09523B

Appendix B: SPARKvue HD Calibration

You can use the SPARKvue HD software if you are using the 
Chemistry MultiMeasure Sensor with a PASPORT-compatible 
interface connected to a computer, or with a mobile device paired 
with a PASCO wireless interface, such as the AirLink2 or 
SPARKlink Air. 

If the sensor is connected to a PASPORT-compatible interface, 
start the SPARKvue software on the computer. If the sensor is con-
nected to a SPARK SLS interface, start the interface. 

For a two-point calibration of the Temperature probe, you will 
need two “known standards,” such as two liquids of known tem-
perature, such as ice water and hot water. A standard thermometer 
will also be needed.

After startup, the screen will show the list of measurements for the 
Chemistry Sensor. 

1.

Touch “Temperature” and then touch “Show”.

The screen will show a Graph display of temperature and 
time. 

2.

Touch the “Experiment Tools” icon (

) in the Graph display 

screen to open the Experiment Tools menu. In the Experiment 
Tools menu, touch “Calibrate Sensor”.

The screen will show “Calibrate Sensor: Select Measure-
ment”. The Sensor (Chemistry), Measurement (Temperature 
(°C)), and Calibration Type (2 point) are already selected. 

3.

Click “Next” to open the “Calibrate Sensor: Enter Values” 
screen.

4.

The Calibration Point 1 “Standard Value” shows 0.0000 °C. 
(This Standard Value can be changed if necessary. Highlight 
the “Standard Value” and enter a new value if needed). For 
this example, put the temperature probe into a “known stan-
dard” of ice water. Wait a few moments. Click “Read From 
Sensor”. 

5.

The Calibration Point 2 “Standard Value” shows 100.0000 °C. 
For this example, put the temperature probe into a “known 
standard” of hot water along with the thermometer. Wait a few 
moments and read the temperature on the thermometer. High-
light the “Standard Value” and enter the temperature from the 
thermometer. Next, click “Read From Sensor” and then click 
“OK”.

The screen of the Graph display returns.

Follow the same procedure to calibrate pH. Use two buffer 
solutions of known pH value, such as pH 4 and pH 7. 

Touch 

here.

Next, touch 

here.

Summary of Contents for PASPort PS-2170

Page 1: ...yurethane Tubing 60 cm 0 125 inch ID 6 Quick release Connectors qty 4 Hose and Coupling Kit 7 Tubing Connectors qty 4 8 Syringe 60 cc 9 Alligator Clip with Insulator qty 2 each Polyurethane Tubing blu...

Page 2: ...RP port BNC port Align the grooves of the BNC connector on the probe with the pins on the BNC port Push the connector onto the port and turn the connector clockwise until it locks in place If you plan...

Page 3: ...easurement for that display Using SPARKvue HD Connect a PASPORT compatible interface to the computer and start the SPARKvue HD software Refer to the documentation that came with your PASPORT interface...

Page 4: ...measurement can be displayed in units of C F or K You can use the included stainless steel probe or a different probe such as the Fast response Temperature Probe PS 2135 or Skin Surface Temperature Pr...

Page 5: ...Electrode Maintenance and Storage Cleaning If the pH electrode becomes contaminated use one of these methods to clean and restore it After any of these procedures rinse the electrode with deionized w...

Page 6: ...tors to connect the tubing to larger diameter tubing or to a rubber stopper with a hole Note Do not allow liquid to enter the pressure port Voltage Use the voltage component of the multi sensor to mea...

Page 7: ...Model No PS 2170 Specifications 7 012 09523B Voltage Range 10 V to 10 V Accuracy 0 1 V Resolution 0 04 mV Voltage Protection up to 240 V Input Resistance 2 M...

Page 8: ...ept Move to the third step 3 The third step in the panel is to choose the type of calibration Two Standards 2 point is the default Click Next 4 The fourth step in the panel is to calibrate the first p...

Page 9: ...he Graph display screen to open the Experiment Tools menu In the Experiment Tools menu touch Calibrate Sensor The screen will show Calibrate Sensor Select Measure ment The Sensor Chemistry Measurement...

Page 10: ...be in distilled water 5 For Calibration Point 2 touch the down arrow to scroll down to the bottom of the screen Place the probe s sensor into the second known standard such as the pH 7 buffer solution...

Page 11: ...s pH Probe into the first known standard pH 4 buf fer solution Wait a few moments and then press F3 H to Read Pt 1 5 Use the down arrow to scroll down to the second calibration point Highlight Pt 2 mV...

Page 12: ...products or service names are or may be trademarks or service marks of and are used to identify products or services of their respective owners For more information visit www pasco com legal Teflon i...

Reviews: