7300w² Monitor Instruction Manual
Navigate to “20mA Offset” option as follows:
Main Menu > Output Config > Output # Config > Output # (M) 20mA Offset
First set the 4mA offset. Adjustment of the 4mA output will also effect the 20mA output, i.e. increasing
the 4mA output by 0.5mA will also add 0.5mA to the 20mA output. Once the 4mA output has been
adjusted as required, the 20mA can then be adjusted if required.
224081IM Issue 22 Issue Date 27/03/2018
`
Page 37 of 46