7
Press Menu on the
keypad to display the
main menu.
The monitor shows
the date and time of
the last sample. Press
OK.
Select Measurement
Config with the arrow
keys. Press OK.
Use.the.arrow.keys.to.
adjust.the.value..Press.
OK.
Choose Sample
Result and press OK.
The sample result
screen displays sensor
info. Press OK.
The monitor displays
a.success.message..
Press OK.
Dissolved Oxygen Sensor - 100% Open Air Calibration
Select Measurement
Config and press the
OK key.
Choose DO
Saturation and press
OK.
Highlight.Calibrate
and press OK.
The monitor displays
sensor info. Press OK.
Calibrate.the.RDO.sensor.using.the.0%.saturation.method.described.below..See.the.
manual for the Partech 750w
2
.Monitor.to.get.more.information.about.calibration..
Suspended Solids Sensor
- Sample
Result
Calibration
Use.Sample.Result.to.adjust.the.value.stored.earlier.with.the.Take.Sample.function..
You must know the value of the calibration solution based on a lab analysis.
Select
Suspended
Solids
with the arrow
keys. Press
OK
.
Calibrate the sensor in
air, but not in a drafty
environment. Press
OK
to continue.
Wait for a saturation
percentage of roughly
90%. Then press OK.