5
Suspended Solids
Sensor - Zero Calibration
Suspended Solids Sensor
- Set Measurement Range (Set Cal)
Sensor Calibration
Use the Set Zero function to store a zero value. Start by immersing the sensor in a
container of clean water. Use a dark container to prevent interference from ambient
light.
.
Use the Set Cal feature to adjust calibration for a known range of solids based on your
application. The set point should represent the maximum expected value for the
measured media. Final Effluent will have a range much lower than Activated Sludge.
Press Menu on the
keypad to display the
main menu.
The Set Zero screen
displays sensor info.
Press OK.
Select Measurement
Config
with the arrow
keys. Press OK.
Choose Set Zero and
press OK.
Select
Suspended
Solids
with the arrow
keys. Press
OK
.
The monitor displays
a value. Wait for it to
stabilize. Then press
OK.
The monitor displays
a zero value for
five seconds and
then shows live
measurements.
Press Menu on the
keypad to display the
main menu.
Select Measurement
Config
with the arrow
keys. Press OK.
Select
Suspended
Solids
with the arrow
keys. Press
OK
.
Choose Set Cal and
press OK.
228538IM-Iss12