Appendix B: Firmware upgrade instructions
V2.03 NitroFlow
®
series
Controller upgrade warning:
Do not cut the power during the upgrade process.
All log data is erased during the upgrade process (also on the SD-card).
When there is an error during the upgrade procedure, try again from the start.
In case of problems contact Parker technical support
Controller upgrade steps:
1. Switch the main Power switch OFF (consult manual)
2. Remove the SD-card (consult manual)
3. Place the latest version of the IO.bin & LCD.bin files in the root directory of the
SD-card.
4. Remove the file FORCE.txt when present.
5. Place the SD-card in the controller.
6. Switch the main Power switch ON.
7. The IO-board is now upgrading (+/- 30sec)
8. The message initialisation pops up (+/- 60sec)
9. The message software needs update 34 will appear.
10. Switch the main switch OFF & ON
11. The display starts flashing (+/- 30sec)
12. The system beeps 2 or 4 times and then starts up.
13. Enter the menu then go to the login screen.
14. Enter 943363 and press enter.
15. The message Clearing flash will appear (+/- 60sec)
16. The message “Turn the unit off” will appear.
17. Switch the main switch OFF & ON
18. The upgrade process is completed.
19. Check the customer software settings.