K3.1.141e Manual NitroFlow HP_EN.doc                           

 

                                - 18 - 

Default settings software parameters 
Menu What Default 

setting 

 Logs 

Interval 60 

min 

 Local settings 

Language English 

 Local settings 

Pressure Bar(g) 

 Local settings 

Flow LPM 

 Local settings 

Purity %O2 

 Alarm settings 

O2 high 

Active: No 
Unit stop: No 
Level: 5% 
Delay: 30 sec 

 Alarm settings 

O2 low 

Active: No 
Unit stop: No 
Level: 0.5% 
Delay: 30 sec 

 Alarm settings 

Inlet pressure high 

Active: No 
Unit stop: No 
Level: 13 bar(g) 
Delay: 30 sec 

 Alarm settings 

Inlet pressure low 

Active: No 
Unit stop: No 
Level: 2 bar(g) 
Delay: 30 sec 

 Alarm settings 

Outlet pressure high 

Active: No 
Unit stop: No 
Level: 10 bar(g) 
Delay: 30 sec 

 Alarm settings 

Outlet pressure low 

Active: No 
Unit stop: No 
Level: 2 bar(g) 
Delay: 30 sec 

 

Summary of Contents for NitroFlow HP

Page 1: ...w HP_EN doc 1 USER MANUAL NitroFlow HP NITROGEN GENERATOR Parker Filtration Separation B V PO Box 258 4870 AG Etten Leur The Netherlands Tel 31 0 76 508 53 00 Fax 31 0 76 508 53 33 E mail pfsinfo park...

Page 2: ...g the customer The contents of this manual can also be changed without prior warning This manual is valid for the NitroFlow HP in its standard version Parker Filtration Separation B V can therefore no...

Page 3: ...PARTS NITROFLOW HP 13 3 4 PROCESS DIAGRAM 14 3 5 PROCESS SCHEME 15 4 TECHNICAL SPECIFICATIONS 16 4 1 GENERAL 16 4 2 PRODUCTION CAPACITY 19 4 2 1 AIR CONSUMPTION 20 4 3 MAINTENANCE KIT 20 4 4 EXPANSION...

Page 4: ...ING THE NITROFLOW HP 43 7 2 START NITROFLOW HP 43 7 3 STOP NITROFLOW HP 43 7 3 1 Alarm messages 44 8 TROUBLESHOOTING 45 8 1 TROUBLESHOOTING LIST 45 9 MAINTENANCE 47 9 1 MAINTENANCE SCHEME 47 9 2 REPLA...

Page 5: ...l void the safety certification of this product If you have any questions or concerns please call your local representative or the technical services department Europe 44 1622 7233 00 USA 1 800 343 40...

Page 6: ...aution shows a danger that can cause damage to the equipment Follow the instructions Electricity High voltage danger of electric shock Warning Risk for death due to suffocation Risk of fire Oxygen enr...

Page 7: ...age Court Hermitage Lane Maidstone Kent ME16 9NT England Tel 44 1622 7233 00 Fax 44 1622 7287 03 1 4 Certificates The NitroFlow HP meets with the following requirements Subject Applicable standard Dir...

Page 8: ...l trained personnel are allowed to work on the NitroFlow HP The user must be aware of hazards related to operating the NitroFlow HP and processes connected to the NitroFlow HP The user is responsible...

Page 9: ...to work on electric and pneumatic equipment are allowed to do the installation maintenance and repairs Unqualified people are not allowed to repair the equipment Refer to 1 6 Lift the NitroFlow HP wit...

Page 10: ...room where the enriched oxygen is ventilated or vent the enriched air outside Only feed the NitroFlow HP with air Keep the air feed to the NitroFlow HP clean and free of organic solvent vapors and oth...

Page 11: ...ecyclable Optimal installation according to instructions and according to good craftsmanship will result in minimal energy consumption and maximal life of your system According to EC regulations elect...

Page 12: ...A F S H2O H2 O2 N2 B C Fig 3 1 Separation principle A Pressurized air inlet B Hollow fibre membrane C Nitrogen outlet F Fast permeation S Slow permeation Ambient air contains nitrogen 78 1 oxygen 20...

Page 13: ...Oxygen sensor B Hollow fiber membrane C Compressed air filters D SD card E Printed circuit board F Active carbon adsorber G Flow control valve for purity adjustment H Flow control valve for sample fl...

Page 14: ...ow HP can either be connected directly to the nitrogen consumer or to a nitrogen reservoir The latter is advised in case of discontinuous or variable peak demands Fig 3 3 Installation lay out NitroFlo...

Page 15: ...valve PI1 2 Pressure indicator F01 Coarse filter F02 Fine filter C1 Carbon adsorber F03 Dust filter QIA h Oxygen analyser FCV1 Flow control valve for purity adjustment FCV3 Flow control valve for sam...

Page 16: ...ized air specifications Max feed pressure 13 bar g 190 psig Compressed air temperature range 10 50 C 50 120 F Pressure dew point 5 C 41 F Residual oil content 3 0 mg m3 Dimensions and connections Dime...

Page 17: ...K3 1 141e Manual NitroFlow HP_EN doc 17 Parts NitroFlow HP 1x NitroFlow HP 1x Manual Table 4 1 General data...

Page 18: ...h Active No Unit stop No Level 5 Delay 30 sec Alarm settings O2 low Active No Unit stop No Level 0 5 Delay 30 sec Alarm settings Inlet pressure high Active No Unit stop No Level 13 bar g Delay 30 sec...

Page 19: ...HP1 159 004032 NitroFlow HP2 159 004033 NitroFlow HP3 159 004034 Table 4 3 Part numbers NitroFlow HP 4 2 Production capacity Type Nominal product capacity Nm hr Nitrogen purity 99 5 99 98 97 96 95 Nit...

Page 20: ...ir use NitroFlow HP Nm3 hr 20 C 1013mbar at nominal conditions Feed pressure at inlet 7 bar g 100 psig Ambient temperature 20 C 68 F Ambient pressure 1013 mbar a 4 3 Maintenance kit Part part number M...

Page 21: ...ions of personnel refer to 2 1 The NitroFlow HP must be transported and stored within a temperature range between 25 C and 50 C 5 2 Define location Install the NitroFlow HP on a fixed location The loc...

Page 22: ...essed air 5 C 41 F 2 System has sufficient capacity 3 System can produce the required pressure take pressure drop into consideration Checklist when connecting to a dedicated stand alone compressor A c...

Page 23: ...er contaminants and cannot be disposed to the sewer as such 5 6 Connecting to mains electricity Warning Do not connect the NitroFlow HP to the mains electricity with a wall plug as to prevent incorrec...

Page 24: ...K3 1 141e Manual NitroFlow HP_EN doc 24 Fig 5 3 Terminals NitroFlow HP Refer to fig 5 4 for location of power cable inlet N L1 PE N L1 PE...

Page 25: ...ompressed inlet air must be within 10 C 50 F and 50 C 120 F refer to table 4 1 Nitrogen outlet Connect nitrogen outlet to the application or to the reservoir refer to fig 3 3 Oxygen outlet The outlet...

Page 26: ...t contacts In and output signals can be connected to the terminal strip on the printed circuit board refer to 3 3 Fig 5 5 Printed circuit board Terminal strip K5 K6 K7 K8 UA1 UA2 ID1 PI1 PI2 TI TT QAI...

Page 27: ...r switch needs to be connected to the digital input ID1 on the printed circuit board Clamp Function Input output signals ID1 Remote start stop Digital input Nominal input current 10 mA Voltage interna...

Page 28: ...by pushing the return to previous screen button 6 2 Configure software parameters 6 2 1 Main screen Access This is the start up screen that automatically appears when the generator is switched on Fun...

Page 29: ...tart if it was stopped less then 3 minutes ago Remaining time is shown on the display When the generator was stopped longer then 3 minutes ago it will start immediately once pressing the start stop bu...

Page 30: ...enus Symbol Menu Access to log on menu refer to 6 2 3 Access to alarm settings menu refer to 6 2 4 Access to pressure switch menu refer to 6 2 5 Access to options menu refer to 6 2 6 Access to local s...

Page 31: ...6 2 6 After selecting PINCODE YES the default pin code 1234 Select PIN CODE CHANGE if you want to enter your personal pin code In case you forgot your personal pin code enter pin code 7833 to return...

Page 32: ...d Correct pin code has to be entered to have access to menus 6 2 4 Alarm menu Access Touch alarm menu button in settings screen Function Set various alarms The alarms that can be set are Screen Alarm...

Page 33: ...w HP_EN doc 33 To activate an alarm touch the active button A You can select NO YES or AUTO RESET When you select YES or AUTO RESET the rest of the data that needs to be set for the specific alarm wil...

Page 34: ...at which the alarm will appear Level screen 3 6 4 6 0 189 PSIG Inlet pressure level at which the alarm will appear Level screen 5 6 6 6 0 11 BAR G Outlet pressure level at which the alarm will appear...

Page 35: ...sound will stop This can only be done if the actual problem was solved 6 2 5 Pressure switch menu Access Touch pressure switch menu button in settings screen Function Activate and set the pressure sw...

Page 36: ...external device has to be a voltage free switch or relay contact Unit can now be started from a remote location by closing the contact switch By opening the contact switch the unit will be stopped If...

Page 37: ...All text in the screen will appear in the chosen language Pressure B BAR G PSIG Pressure indications will appear in the chosen setting If BAR G is chosen then temperature will be automatically in C I...

Page 38: ...onsists of 5 different screens Each screen displays maintenance status or calibration buttons Data Explanation Type Shows type of generator Version Software version of system O2 lifetime Month year wh...

Page 39: ...her BAR G or PSIG C Inlet Compressed air inlet temperature in Celcius or Fahrenheit When selected BAR G temperature is displayed in Celcius When selected PSIG temperature is shown in Fahrenheit Refer...

Page 40: ...flow by entering flow value measured with external flow meter Replace Filter When a filter has been replaced during maintenance button can be touched and countdown for the new filter is set System ask...

Page 41: ...nly alarms Alarms as well as status of the sensors temperature pressure O2 flow and maintenance are saved on the SD card The time between the logging saving of this data on the memory card can be chos...

Page 42: ...r for this software Your PC also needs a SD card reader 1 Switch off the generator 2 Remove the SD card refer to 3 3 3 Refer to appendix A for information how to use the NitroView software 4 Replace t...

Page 43: ...vel and vice versa The oxygen level can be read on the display refer to 3 3 7 Check whether the system works according to the settings of the control system 8 Close all sheet metal covers Important Th...

Page 44: ...utlet high Outlet pressure too high Off P outlet low Outlet pressure too low Off T inlet high Inlet temperature too high On T inlet low Inlet temperature too low On P Inlet sensor fail Inlet pressure...

Page 45: ...ting Sample flow to O2 sensor is too low Check readjust the sample flow to the O2 sensor Refer to 9 5 Nitrogen outlet line is restricted Check open the outlet line Delivery flow of nitrogen too low Fa...

Page 46: ...low to the O2 sensor Refer to 9 5 O2 sensor drifted Calibrate O2 sensor Refer to 9 5 Required O2 concentration can not be reached O2 sensor used up Replace O2 sensor Refer to 9 4 Switch off pressure s...

Page 47: ...8000 running hours which comes first Active carbon adsorber Replace adsorber 1x per year together with filter elements Check purity setting Determine oxygen content in output and readjust by readjust...

Page 48: ...at the filter is depressurized 6 Turn the filter bowl D clockwise and pull the bowl from the filter housing A 7 Unscrew the blue knob C 8 Remove the old filter element B 9 Place a new filter element B...

Page 49: ...leakages after restarting NitroFlow HP 9 4 Replace oxygen sensor 1 Stop and switch off the NitroFlow HP 2 Remove the cover plate above inlet and outlet connections of the NitroFlow HP 3 Disconnect co...

Page 50: ...wait until the oxygen level has stabilized this will take approx 60 seconds 5 Go to screen 5 5 and touch O2 20 9 span button The system will ask for confirmation select YES 6 Start the NitroFlow HP b...

Page 51: ...1 Stop the unit by touching the button in the main screen refer to 6 2 1 2 Switch off the main switch 3 Remove the cover plate above inlet and outlet connections 4 Remove the fuse holder A 5 Place a...

Page 52: ...K3 1 141e Manual NitroFlow HP_EN doc 52 10 Electrical scheme and terminal connections Fig 10 1 Electrical scheme NitroFlow HP...

Page 53: ...K3 1 141e Manual NitroFlow HP_EN doc 53 Fig 10 2 Electrical terminal connections NitroFlow HP...

Page 54: ...et filter 48 Installation 21 L Liability 8 Local settings menu 37 Location 21 Log on menu 31 M Main screen 28 Maintenance 47 Maintenance kit 20 Maintenance menu 38 Menu structure 28 N Nitrogen 9 nitro...

Page 55: ...be found on this bar There are 2 menus File main menu and about 2 System info Info about the log you are viewing Info about the hardware the log came from 3 View options Enable extra columns of the lo...

Page 56: ...the type of log see Preferences When selecting a line the alarms will be showed on the bottom of the screen 8 8 Alarm bar This bar shows alarm states of the selected log When you ve selected a log th...

Page 57: ...or as well For example if you use dark colors for the events you can choose white text Just click on a color box then you can pick a color 3 Default colors Click this to reset all colors back to defau...

Page 58: ...4 Remove the file FORCE txt when present 5 Place the SD card in the controller 6 Switch the main Power switch ON 7 The IO board is now upgrading 30sec 8 The message initialisation pops up 60sec 9 The...

Page 59: ...nu functionality improved Compressor hour counter reset code added V2 0 03 December 2007 New version because of hardware changes Service messages no buzzer Filter lifetime counter adjusted Data log me...

Reviews: