– Temperature: 132 °C / 134 °C
– Holding time: 3 minutes
PROCEDURE:
• Pack all of the disassembled parts in a sterile barrier system
in accordance with DIN EN 11607-1 (e.g., foil-paper pack-
aging).
• Carry out the sterilisation in a steam steriliser in accordance
with the manufacturer’s instructions.
Sterilisation temperature and holding time:
132 ºC / 134 °C, at least 3 minutes.
DRY:
Ensure that there is no residual moisture remaining in the com-
ponents. Optionally:
• Shake the water out of all of the parts.
• Place all parts on a dry, clean and absorbent surface and al-
low them to dry completely.
Visual inspection and storage
Storage location:
– dry
– dust-free
– protected from sources of contamination
optional: Use sterile packaging
– 30 –