Chemical
cleaning with
disinfection:
EQUIPMENT:
The method has been validated in Europe using:
– Aldehyde-free instrument disinfectant: Bode
Bomix
®
plus (concentration: 2%)
Active agent basis: Quaternary ammonium
compound
– Drinking water at about 15 °C
– Application time: 5 minutes
PROCEDURE:
•
Clean and disinfect the individual parts in a
single work step, using a solution prepared
according to the manufacturer’s instructions.
Info:
If the recommended treatment time is
exceeded significantly, the plastic parts may
take on the smell of the disinfectant.
RINSING:
•
CAUTION!
Disinfectant residues can
cause allergic reactions or irritation of the mu-
cous membrane. Rinse off all parts thoroughly
in running drinking water at about 15 °C for 3
minutes.
•
Dispose of the used solution (the diluted solu-
tion can be disposed of down the sink).
DRYING:
•
Shake the water out of all of the parts.
•
Place all parts on a dry, clean, and absorbent
surface, and allow them to dry completely.