9
en
Cleaning Aid für eFlow
®
Aerosol Head
Part no. 078G6100
Read these instructions
carefully and completely
before using
for the fi rst time. Keep these instructions in a safe place.
1
Product description
Hinged cleaning aid for cleaning the eFlow
®
aerosol head with a built-in fun-
nel opening and receptacle for fi lling 5 ml isotonic sodium chloride solution
(0.9%). The nebuliser connection cord and the control unit of the eFlow
®
appliance are also required for operation. Information on connecting the
control unit to the power supply and on operation can be found in the Instruc-
tions for Use of the eFlow
®
appliance.
2
Intended use
The easycare cleaning aid can contribute to a shortening
of the nebulisation time if the performance of the eFlow
®
aerosol head is deteriorating.
As a result of easycare cleaning, the pores in the membrane
of the eFlow
®
aerosol head are mechanically rinsed free by
means of backwashing in the opposite direction.
3
Important information for use
Cleaning with the easycare cleaning aid is recommended if the nebulisation
time for the same fi ll volume of a given substance signifi cantly increases.
Additional information can be found in the current version of the corre-
sponding eFlow
®
Instructions for Use in the Chapter “Troubleshooting”, key
word “prolonged nebulisation times” (www.pari.de/Products/Pari Inhalation
devices/eFlow).