10
Example eFlow
®
rapid
•
Measure the nebulisation time with 2.5 ml isotonic sodium chloride solu-
tion (0.9%) without inhaling.
•
If the nebulisation time takes longer than
4 minutes
,
cleaning with the easycare cleaning aid is recom-
mended.
Frequency of use
If needed, no more than 2 times weekly after inhalation and after cleaning
the aerosol head (see cleaning steps below).
Note:
More frequent use is
not
beneficial, but can lead to premature material
wear.
During its entire period of use, an aerosol head should not be back-
washed more than 25 times with the easycare cleaning aid.
The easycare cleaning aid does not extend the life cycle of the aerosol
head.
Backwash liquid
5 ml isotonic sodium chloride solution (0.9%)
Cleaning steps
The easycare backwashing procedure is performed as an
additional
meas-
ure for the normal cleaning of the aerosol head.
Please ensure that the following cleaning steps are observed
during easycare backwashing:
1. Cleaning nebuliser parts and aerosol head *
)
2.
Backwashing the aerosol head with easycare
3. Disinfecting/sterilising nebuliser parts and aerosol head *
)
*) Points 1 and 3 see Chapter “Hygienic Re-use” in the eFlow
®
Instruc-
tions for Use