Changing the filter

Under normal operating conditions the filter in the front panel of your PARI BOY 
N compressor must be replaced every 500 hours of operation or at least once a 
year.
In addition, the filter must be inspected at regular intervals (after every 10th to 
12th inhalation). If it is soiled (grey or brown staining) or blocked it should be 
replaced immediately. If the filter has become moist it should likewise be ex-
changed for a new one.
Do not attempt to clean the filter and reuse it!
Use only the original PARI filter. Your PARI compressor could otherwise be dam-
aged or the quality of the therapy may be impaired.

Replace the filter as follows:
• Remove the filter holder from your PARI BOY N by unscrewing it in an anti-

clockwise direction with a suitable coin.

• Replace the filter and insert the filter holder into your PARI BOY N. Screw it 

tight with a coin by turning in a clockwise direction.

PBoy.book  Seite 16  Mittwoch, 25. September 2002  9:53 09

Summary of Contents for BOY N

Page 1: ...BOY N Kompressor PBoy book Seite 1 Mittwoch 25 September 2002 9 53 09 2001 PARI GmbH Spezialisten f r effektive Inhalation 085D1001 9 02...

Page 2: ...2 PBoy book Seite 2 Mittwoch 25 September 2002 9 53 09...

Page 3: ...and for topical inhalation in anaesthesia The PARI BOY N compressor can be used with all PARI nebulisers except PARI LC PLUS UNIVER SAL Please ensure that all the following components are included wit...

Page 4: ...Information 1 2 3 4 5 1 ON OFF switch 2 Connecting tube 3 Filter 4 Holder for nebuliser 5 Power socket 6 Rating plate underneath the compressor 6 PBoy_GB fm Seite 12 Mittwoch 25 September 2002 12 14...

Page 5: ...nd to the PARI LC extension ring The nebuliser can then easily be attached to the LC extension ring The con necting tube may remain connected to the LC extension ring for easier han dling Make sure th...

Page 6: ...or by pressing the ON OFF switch Return the nebuliser to its holder after the treatment has been completed Warning To avoid health risks e g infection please read the Hygiene sec tion in the instructi...

Page 7: ...nvironmental conditions condensation may collect in the neb uliser connecting tube during inhalation The moisture must be removed to guarantee optimum treatment Please proceed as follows Disconnect th...

Page 8: ...ilter has become moist it should likewise be ex changed for a new one Do not attempt to clean the filter and reuse it Use only the original PARI filter Your PARI compressor could otherwise be dam aged...

Page 9: ...ppliance It should not be operated without supervision or in areas with a risk of explosion or high humidity Thor oughly acquaint yourself with these instructions for use before using the appli ance a...

Page 10: ...from faults in ma terial or workmanship The owner shall not be entitled to cancel the sale or de mand a partial or full refund of the purchase price Damage caused by improper use of the appliance is n...

Page 11: ...hin the guarantee period we or an authorised customer service point will repair free of charge any defects in the appliance resulting from faults in material or workmanship Damage arising from imprope...

Reviews: