4.3 Cleaning
• Place all disassembled com-
ponents in warm tap water with
a little dishwashing liquid for at
least 5 min. If necessary, use a
clean brush to remove loose
dirt (the brush must be re-
served exclusively for this pur-
pose).
• Rinse all parts thoroughly in running water.
• You can remove excess water more quickly by shaking all
parts.
4.4 Disinfection
After cleaning, disinfect the dismantled nebuliser and accessor-
ies used (except for the connection tubing) (effective disinfec-
tion is only possible if the nebuliser has been cleaned).
Information when using masks with rubber bands:
The rubber band can only be cleaned as it might be dam-
aged by the high temperatures used during disinfection.
CAUTION
A damp environment may encourage the growth of bacteria.
Therefore, remove all parts from the pot or disinfector as
soon as disinfection has finished. Dry the parts. The
risk of
infection
is reduced when the parts are dried completely.
– 52 –