21.2 Compliance level
The interference immunity levels in accordance with IEC 60601 are fulfilled.
Manufacturer's declaration - electromagnetic interference
Biolution is designed to be operated under the electromagnetic conditions specified below. The customer or
the user of the device should ensure that it is used under these conditions.
Emission measurements
Compliance
Electromagnetic conditions -
guidelines
HF emissions in accordance with
CISPR11
Group 1
Biolution only uses HF energy for
its internal functions. Therefore its
HF emissions are very low, and it
is unlikely that adjacent electronic
devices will be disturbed by these.
HF emissions in accordance with
CISPR11
Class B
The device is designed to be used in
all set-ups including living areas and
those which are directly connected
to a public power grid that also
supplies power to buildings being
used for living purposes.
Emission of harmonics in accordance
with IEC61000-3-2
Class A
Emission of voltage fluctuations/
flickers in accordance with IEC61000-
3-3
fulfilled
Table 201 in accordance with DIN EN 60601-1-2, 6.8.3.201