Customer Guide
HPAC Cell Cleaning
©2015 Parata Systems, LLC
page 32 of 40
Parata Technical Assistance Center
120-0465 Rev 03
1-888-PARATA1 (727-2821)
7.5.b
Carefully insert the foam plug in the opening next to the black height
setting knob on the bottom of the cell.
7.6
Obtain a nozzle splash guard from the cleaning kit and install it.
7.6.a
Peel the backing from the splash guard.
7.6.b
Align the opening in the splash guard with the nozzle opening, place
the guard on the nozzle, and apply pressure to secure it in place.
8.
Follow the replenishment procedure in the Mini Operator’s Guide to load pills into the cell.
•
If you are placing the same drug in the cell:
•
Follow standard procedures to calibrate the cell.
Note
You must calibrate the cell even if you are placing the same drug in the cell.
The settings will have changed during the cleaning process.
•
Follow the replenishment procedure in the Mini Operator’s Guide to load pills into
the cell.
•
If you accidentally deleted the cell from the system:
•
Follow standard procedures to calibrate the cell.
Note
You must calibrate the cell even if you are placing the same drug in the cell.
The settings will have changed during the cleaning process.
•
Follow the process in the “Setting up a new cell” section in the Mini Operator’s
Guide to assign a drug to the cell.
•
If you are putting a new drug in the cell:
•
Follow the process in the “Deleting a cell” section in the Mini Operator’s Guide to
remove the cell from the system.
•
Then follow the process in the “Setting up a new cell” section in the Mini
Operator’s Guide to configure the cell for the new drug.
CAUTION
Do not
add pills to the cell until the cell is completely dry.