Function
Range
Resolution
Accuracy of the
value displayed in %
*)
DC voltage (V=)
(min. 0,5 V=)
600 V
0, 1 V
±(0,8% + 3 digits
AC voltage (V~)
50 to 60 Hz
(min. 1,0 V~)
600 V
0,1 V
±(1,2% + 5 digits)
Resistance (Ω)
10 M
1
±(1,2% + 3 digits)
Frequency (Hz)
(min. 1,0 V~)
10 to 1000 Hz
1 Hz
±(2,0% + 10 digits)
Capacitance
*)
(F)
(min. 0,5 pF)
1000 µF
0,1 µF
±(3,5% + 8 digits)
6000 µF
1 µF
±(4,5% + 10 digits)
*) Overload protection Capacity measurement: max. 250 V DC or AC (RMS)
Function
Range
Diode testing
1 V
Automatic detection in the event of a
voltage drop of <1,0 V
Audible continuity
<30
If there is a resistor <30 Ω, you hear a
beep. When the circuit is open, the
display shows "OL".
Non Contact Voltage
tester
low
Note chapter 8 Operating Instructions,
Non Contact Voltage tester
medium
high
Accuracy is valid for one year from the last calibration.