Part names and functions
9
1. Connect the sensor to UV CHK.
2. Align the center of the LED head with the center of the light-
receiving section of the UV sensor.
3. Hold down .
(The AUTO mode display switches to the TYPE00 setting mode display.)
4. Press .
5. Press .
((The selection of CH1 will be accepted, and irradiation will be ready
(if the sensor is connected).)
6. Press .
(UV irradiation will start, and the measured value will be indicated on the
display.)
7. Press .
(Irradiation will stop.)
1. Follow procedures 6-(1) to (4) to measure the UV intensity, and
then calibrate it by pressing and .
(The calibrated UV intensity will be displayed.)
2. Press .
(To accept the calibrated UV intensity level and stop irradiation)
3. Hold down .
(To save the UV intensity level of CH1 in memory and go back to AUTO
mode)
Switch to setting mode.
Prepare for the measurement.
Measure the UV intensity and
calibrate it.
End the calibration.
Select CH1.
Measure the UV intensity.
How to measure the UV intensity (e.g. CH1)
End the measurement.
You can check the UV intensity of the simple mode program by using the optional UV sensor.
How to manually calibrate the UV intensity (e.g. CH1)
You can calibrate the UV intensity set for the simple mode program based on the measured intensity.