  ALARMS & SAFETY FUNCTIONS 

 
This unit has the alarms and safety functions shown in Table 6, and also self diagnostic functions. 
 

Table 6    Alarms and safety functions 

Alarm & Safety 

Situation 

Indication 

Buzzer 

Safety operation 

High temperature 
alarm 

If the chamber temperature is higher 
than the temperature at which the 
high temperature alarm is activated. 

Alarm lamp is flashed. 
Digital temperature indicator is
flashed. 

Intermittent tone with 
15 minutes delay. 

Remote alarm with 15 
minutes delay. 

Low temperature 
alarm 

If the chamber temperature is lower 
than the temperature at which the 
low temperature alarm is activated. 

Alarm lamp is flashed. 
Digital temperature indicator is
flashed. 

Intermittent tone with 
15 minutes delay. 

Remote alarm with 15 
minutes delay. 

Power failure alarm 

In the case of power failure. 
When power switch is turned OFF. 
When the power to the unit is 
disconnected. 

Alarm lamp is flashed. 

Intermittent tone 

Remote alarm. 

Auto-return 

When there is no key pressing in 
each setting mode for 90 seconds. 

Chamber temperature is 
displayed. 

----- 

 

Finishing of each 
setting mode. 

Key lock 

When the key lock is “ON”. 

----- 

----- 

Change of setting is 
disable. 

Battery switch 

check 

When battery switch is OFF at the 
time of alarm test. 

Alarm lamp is flashed. 
E09 is flashed. 

Intermittent tone 

Remote alarm. 

Thermal sensor 
Abnormality 

If the thermal sensor is 
disconnected. 

Alarm lamp is flashed. 
E01 and chamber temp. 
are displayed alternately. 

Intermittent tone 

Remote alarm. 
Continuous running. 

If the thermal sensor is 
short-circuited. 

Alarm lamp is flashed. 
E02 and chamber temp. 
are displayed alternately. 

Intermittent tone 

Remote alarm. 
Continuous running. 

Ambient sensor 
abnormality 

If the ambient temperature sensor is 
disconnected. 

Alarm lamp is flashed. 
E03 and chamber temp. 
are displayed alternately. 

Intermittent tone 

Remote alarm. 

If the ambient temperature sensor is 
short-circuited. 

Alarm lamp is flashed. 
E04 and chamber temp 
are displayed alternately. 

Intermittent tone 

Remote alarm. 

Compressor 
protection sensor 
abnormality 

If the compressor protection   
temperature sensor is disconnected. 

Alarm lamp is flashed. 
E05 and chamber temp.   
Are displayed alternately. 

Intermittent tone 

Remote alarm. 
Normal operation. 

If the compressor protection 
temperature sensor is short 
circuited. 

Alarm lamp is flashed. 
E06 and chamber temp. 
Are displayed alternately. 

Compressor temp. 
Abnormality 

In the case of failure of compressor
cooling fan motor. 
In the case of abnormal high temp. 
due to the dust on the condenser. 
In the case of abnormal high 
ambient temperature. 

E10 and chamber temp. are 
displayed alternately. 

Intermittent tone 

Remote alarm. 
Compressor stops 
running when the temp
reaches about 80

o

C. 

Battery check 

When about three years has passed 
with power switch ON. 

_F1 and chamber temp. 
are displayed alternately. 

----- Information 

only 

Fan motor check 

When about six years has passed 
with power switch ON. 

_F2 and chamber temp. 
are displayed alternately. 

----- Information 

only 

 

 

  The above power failure alarm is available when the battery switch is ON and the battery is charged.    If 

the battery switch is OFF or the battery is discharged, only the remote alarm is activated. 

 

 The power failure alarm can be kept about twelve hours with the battery charged completely.  2-day 

operation of the freezer is needed to charge the battery full. 

 

 The chamber temperature is displayed for five seconds if the alarm buzzer stop key (BUZZER) is 

depressed during the power failure alarm.    At the same time, the alarm buzzer stops. 

 

 The remote alarm is stopped by pressing alarm buzzer stop key (BUZZER) as the remote alarm is 

operated in conjunction with the buzzer, except for the power failure alarm. 

 

 After power failure, the operation is resumed with the condition before power failure since the 

temperature setting and alarm temperature setting are memorized in a nonvolatile memory. 

 

  The following directions are the information of the part replacement (it is recommendation).    This is not 

a trouble. 

a) The battery for power failure alarm is an article for consumption.    It is recommended that the battery 

will be replaced about every three years.  Contact our sales representative or agent at the time of 
replacement of the battery for recycling. 

b) Fan motors are expendable supplies.  Replace it for about every six years.  Contact our sales 

representative or agent at the time of replacement of the fan motor. 

Summary of Contents for MDF-U334 Series

Page 1: ...lease read these instructions carefully before using this product and save this operating instructions for future use See page 33 for all Model Numbers MDF U537 U537D MDF U334 Biomedical Freezer Opera...

Page 2: ...ALARM TEMPERATURE SETTING P 15 SETTING OF ALARM RESUME TIME P 17 ALARMS SAFETY FUNCTIONS P 18 ROUTINE MAINTENANCE P 19 Cleaning of cabinet P 19 Defrosting P 19 TROUBLESHOOTING P 20 DISPOSAL OF UNIT P...

Page 3: ...ctions will be subjected to change without notice due to the improvement of performance or functions Contact our sales representative or agent if any page of the operating instructions is lost or page...

Page 4: ...s caution this symbol means an action is prohibited this symbol means an instruction must be followed Be sure to keep this operating instructions in a place accessible to users of this unit Some warni...

Page 5: ...wer supply outlet is not grounded it will be necessary to install a ground by qualified engineers Never ground the unit through a gas pipe water main telephone line or lightning rod Such grounding may...

Page 6: ...nction Disconnect the power supply plug if there is something wrong with the unit Continued abnormal operation may cause electric shock or fire When removing the plug from the power supply outlet grip...

Page 7: ...cause a heat or ignition Never store corrosive substances such as acid or alkali in this unit if the container cannot be sealed These may cause corrosion of inner components or electric parts Check t...

Page 8: ...40 o C Maximum relative humidity 80 for temperature up to 31 o C decreasing linearly to 50 relative humidity at 40 o C Mains supply voltage fluctuations up to 10 of the nominal voltage Transient overv...

Page 9: ...8 FREEZER COMPONENTS 2 11 3 9 8 7 6 4 1 5 12 Hole of Padlock Back side 15 14 13 N C COM N O 10...

Page 10: ...areful not to damage the container by a metal scraper at the time of defrosting WARNING Fix the storage container securely Incomplete installation may cause injury or damage Never touch the storage it...

Page 11: ...s flashed Pressing this key again after setting desired temperature the setting is stored into computer memory If there is no key operation for ninety seconds during the temperature setting mode the t...

Page 12: ...y cause vibration or noise WARNING Install the unit on a sturdy floor If the floor is not strong enough or the installation site is not adequate this may result in injury from the unit falling or tipp...

Page 13: ...contact them to the floor Ensure the unit is level 3 Fixing the unit Two fixtures are attached to the rear of the frame Fix the frame to the wall with these fixtures and rope or chain 4 Ground earth...

Page 14: ...SETTING Table 1 shows the basic procedure for setting the chamber temperature Perform key operations in the sequence indicated in the table The example in the table is based on the assumption that the...

Page 15: ...ent chamber temperature is displayed The key lock only works with the chamber temperature setting and the defrost key DEF REMOTE ALARM TERMINAL WARNING Always disconnect the power supply cord before c...

Page 16: ...alarm can activate when the chamber temperature is 5 o C lower than the set temperature Table 3 Procedure for setting high temperature alarm Description of operation Key operated Indication after oper...

Page 17: ...of settable digit increases 4 Press set key SET The right digit is flashed 5 Set the temperature to 05 with the digit shift key and numerical value shift key Pressing the key shifts the digit which c...

Page 18: ...3 Set the figure to F25 with the digit shift key and numerical value shift key Pressing the key shifts the digit which can be set When pressed the figure of settable digit increases 4 Press set key S...

Page 19: ...cted Alarm lamp is flashed E05 and chamber temp Are displayed alternately Intermittent tone Remote alarm Normal operation If the compressor protection temperature sensor is short circuited Alarm lamp...

Page 20: ...nical parts are completely sealed This unit requires absolutely no lubrication Defrosting Defrost the freezer whenever one of the defrost marks shown by 12 in the figure on page 9 is hidden by frost U...

Page 21: ...sunlight There is any heating source near the freezer A rubber cap and insulation for the access port are not set correctly You put too many unfrozen articles into the freezer compartment When the uni...

Page 22: ...s remove doors to prevent accidents such as suffocation Recycle of battery The unit contains a rechargeable battery The battery is recyclable At the end of it s useful life check with you local solid...

Page 23: ...pose of them correctly at your local community waste collection recycling centre Please help us to conserve the environment we live in German F r EU Staaten Das Symbol und das erw hnte Wiederverwertun...

Page 24: ...es de mani re inappropri e peuvent s av rer potentiellement dangereuses pour la sant humaine et pour l environnement Cependant ces mati res sont n cessaires au bon fonctionnement de votre appareil ou...

Page 25: ...m recolha de lixo local Por favor ajude nos a conservar o ambiente Italian PER UTENTI UE Il simbolo e i sistemi di riciclaggio descritti di seguito si applicano esclusivamente ai paesi dell UE Questo...

Page 26: ...en bij uw plaatselijk afvalinzameling recyclingcentrum in uw buurt Help ons het milieu waarin wij leven in stand te houden Swedish F R ANV NDARE INOM EU Den symbolm rkning och de tervinningssystem som...

Page 27: ...and disconnect the power supply plug 2 Remove six screws fixing the top cover with a screw driver and remove the top cover Fig 1 3 Remove two screws fixing the battery mounting plate Fig 2 and discon...

Page 28: ...l freezer as the optional component The type of the temperature recorder is MTR G85A MTR G85C and MTR 4015LH For the attachment the recorder fixing is necessary Following shows the combination of reco...

Page 29: ...b counterclockwise See Fig 2 3 As shown in Fig 3 insert the recording paper with the begin tab placed in the cartridge Check that the printed side is facing out 4 Place the recording paper beneath the...

Page 30: ...paper has been fed into the used recording paper compartment send all the recording paper in the compartment first turning the gear Battery replacement To replace the battery turn the knob countercloc...

Page 31: ...rows to open the head clamp at A and B See to Fig 2 illustration 1 4 Position the ink pen so that the guide pins fit into the guide holes on the pen arm See to Fig 2 illustration 2 5 Press the two sid...

Page 32: ...nter of the recording paper is divided into thirty two equal sections The lines extending from these lines serve as the thirty two day time scale Fig 4 Stopping recording 1 Turn OFF the power switch 2...

Page 33: ...8 o C and 35 o C Temperature display Digital display between 50 o C and 50 o C Temperature sensor Thermistor sensor Temperature alarm Flash of digital indicator and alarm lamp Buzzer Remote alarm Acce...

Page 34: ...ature 35 o C no load Temperature control range 20 o C to 30 o C Rated voltage AC 110 V AC 115 V AC 220 V AC 220 V AC 220 V 230 V 240 V Rated frequency 60 Hz 60 Hz 50 Hz 60 Hz 50 Hz Rated power consump...

Page 35: ...he unit Unit interior No contamination Decontaminated Contaminated Others 3 Instructions for safe repair maintenance of the unit a The unit is safe to work on b There is some danger see below Procedur...

Page 36: ...Printed in Japan 7FB6P151619001 S0412 10412 1 1 1 Sakata Oizumi Machi Ora Gun Gunma 370 0596 Japan Panasonic Healthcare Co Ltd 2012...

Reviews: