CHAPTER 5 SPECIFICATIONS
5-3
Laser
emission
indicator
Green LED (Lights up during or immediately before emission of
laser from sensor head 1 or 2)
BRIGHT
indicator
Red LED (Lights up upon measurement failure of sensor head
1 or 2 due to excessive light amount)
12
Indicato
rs
DARK
indicator
Red LED (Lights up upon measurement failure of sensor head
1 or 2 due to insufficient light amount)
13
No. of connected
sensor heads
2 max.
14
Setting / data display
Compact console (optional) (*2)
Shift
200.0000mm
15
Calibration
(
3)
Span
0.9000 to 1.1000
16
Average
frequency ( 3)
OFF, 2 to 32, 768 times (16 steps)
17
Digital filter ( 3)
High pass: OFF, 10 to 2,000 Hz (9 steps)
Low pass: OFF, 10 to 2,000 Hz (9 steps)
18
Calculation
function ( 3)
L KA , L KB , L K
(
A B
)
A, B: Measurement of sensor head 1 or 2
L= 999.9999 , K=0.0001 to 99.9999
19
Hold function
( 3)
Normal / P-P / PEAK / VALLEY
20
Vibration
resistance
10 to 55Hz frequency (period: 1 min.), 0.75mm amplitude in X,
Y and Z directions for 30 min. each
21 Shock
resistance
196m/S
2
in X, Y and Z directions for three times each
22
Ambient
temperature
0 to +50
℃
(No dew condensation or icing allowed)
Storage: -20 to +70
℃
23
Ambient humidity
35 to 85% RH
24 Material
Enclosure:
Aluminum
25 Weight
300g
approx.
Note: The measuring conditions are 24VDC source voltage, 20
℃
ambient temperature,
100 s sampling frequency, 512 time average frequency and measurement center
distance unless otherwise specified.
1 : The maximum output current of the analog current output varies according to the load
impedance.
2 :
For details of the optional compact console [
HL-C1DP1
(
-E
)
-WL
], refer to the
specifications of the compact console.
3 : Data can be set with the compact console or command inputs from an external device
via RS-232C.
Summary of Contents for HL-C1-WL Series
Page 4: ...2 English Japanese Measuring range indicator Laser emission indicator Warning label ...
Page 11: ...CHAPTER 1 NAME AND FUNCTION OF EACH PART ...
Page 14: ...CHAPTER 1 NAME AND FUNCTION OF EACH PART 1 3 MEMO ...
Page 15: ...CHAPTER 2 INSTALLATION OF SENSOR HEAD ...
Page 16: ...CHAPTER 2 INSTALLATION OF SENSOR HEAD 2 1 2 1 Mounting Dimensions Hl C135C BK10 Unit mm ...
Page 19: ...CHAPTER 3 INPUT TERMINAL BLOCK ...
Page 27: ...CHAPTER 5 SPECIFICATIONS ...
Page 33: ...CHAPTER 6 DIMENSIONS ...