Figure 3-2: Unknown Sound Material Do Screen
Step 1:
Update the Transducer Zero Compensation by wiping the
transducer face clean of all couplant and pressing [2
nd
F],
[Cal Zero](Do Zero).
Step 2:
Couple the transducer to the block.
Step 3:
Press the [CAL VEL] key.
Step 4:
When the thickness reading is stable, press the [ENTER]
key.
Step 5:
Uncouple the transducer and use the [
,
,
,
and ]
slewing keys to enter the thickness of the standard.
Step 6:
Press the [MEAS] key to complete the calibration and return
to Measure mode.