910-257C
March 2006
Page 9-4
MG2 Series
material itself probably cannot be measured due to:
• Extreme near side or far side surface roughness
• Extremely high sound attenuation or scattering due to
graininess, inclusions, voids or, other material
properties
• Extreme non-parallelism
• Excessively sharp curvature
9.6 Self Diagnostics
MG2 Series corrosion thickness gages include two self-diagnostic
screens that permit the user to identify hardware or software
problems.
To view the Diagnostic 1 screen and see internal self-test results:
Step 1:
Press [2ndF][
](Setup) to display the setup tabs.
Step 2:
Use the [
,
] keys to highlight the DIAG1 tab.
The following results are displayed: