3.3 Velocity and Zero
Calibration
The Material Velocity and Zero Calibration procedures may be
combined using a thick and a thin calibration block of the same
material.
Step 1:
Update the Transducer Zero Compensation by wiping the
transducer face and pressing [2
nd
F], [Cal Zero](Do Zero).
Step 2:
Couple the transducer to the thick calibration block.
Step 3:
Press [CAL VEL].
Step 4:
When the thickness reading is stable, press the [ENTER]
key.
Step 5:
Uncouple the transducer from the block and use the [
,
,
,
and
] slewing keys to enter the thickness of the
thick block.
Step 6:
Couple the transducer to the thin block and press the [CAL
ZERO] key.
Step 7:
When the reading is stable, press the [ENTER] key.
Step 8:
Uncouple the transducer from the block and use the [
,
,
,
and
] slewing keys to enter the thickness of the thin
block.
Step 9:
Press the [MEAS] key to complete the calibration and go to
the Measure mode.