The following symbols may appear in this manual, on the Digital Blood
Pressure Monitor B63, or on it's accessories. Some of the symbols represent
standards and compliances associated with the Digital Blood Pressure
Monitor B63 and its use.
5.5. Error Indicates
The following symbol will appear on the display when measuring abnormal
SYMBOL DESCRIPTIONS
ER1
ER2
ER3
ER4
ER6
ER7
ER8
SYMBOL
No display
appears
Is the battery low
?
Sensor abnormal
Please charge in time
CAUSE
CORRECTION
Check if the pump is working or not, if it
is working, then the problem is sensor
abnormal, Please contact the
manufacturer
Re measure, Please contact the
manufacturer
Re measure, Please contact the
manufacturer
Tie the cuff and measure again
Re measure, Please contact the
manufacturer
Please contact the manufacturer
Please contact the manufacturer
Not enough heartbeat detected
or blood pressure calculated
Abnormal measurement results
The cuff is too loose
Pressure fluctuation is large when
measuring
The pressure exceeds the upper
limit
Abnormal or uncalibrated
calibration data
Authorized Representative in the European Community
CE Mark: conforms to essential requirements of the Medical
Device Directive 93/42/EEC.
Date of manufacture.
Manufacturer
Specifies serial number
Type BF applied part
Direct current
DISPOSAL: Do not dispose this product as unsorted municipal
waste. Collection of such waste separately for special treatment
is necessary.
Follow instructions for use
Trouble removal
Problem
Please contact the dealer if you cannot solve the problem, do not disassemble the
device yourself!
(
Manufacturer's Email: [email protected]
)
check
cause and solutions
No power
check the battery power
Please charge in time
Err and stop working whether move the arm when inflate
keep the body peaceful
cuff leak
whether the cuff wrap too loose
whether the cuff broken
wrap the cuff tightly
Please contact the
manufacturer
check if chatting when measured
keep quite when measure