Insert the lowest turbidity
value calibration standard
when prompted.
Use the Up/Down buttons
to adjust the value on
screen to that on the
standard and press ‘Read.
The instrument will request
the remaining standards
in order of increasing
turbidity value. After the
final standard has been
read, the screen will show
a countdown timer. When
the calibration process is
complete the screen will
show ‘Successful’.
Turbimeter
Formazin
Read
and press Read
System
10.0
01/01/2017 09:30:30
User 1
123456
Back
and press Read
Insert Standard
3
Continued...
27
The time, date and Operator ID for successful
calibrations are stored in the calibration log.
To manage the calibration log see page 38.
Turbimeter
Formazin
Read
and press Read
System
0.40
01/01/2017 09:30:30
User 1
123456
Back
and press Read
Insert Standard
2