Turbimeter
Check Standard Mode
OK
Back
01/01/2017 09:30:30
User 1
123456
Check Standard Read
Check Standard Values
Turbimeter
1.00 NTU
Read
Insert Standard
Press ‘Read’
01/01/2017 09:30:30
User 1
123456
Cancel
Select the Check
Standard Read option
to perform a recorded
reading of each turbidity
check standard.
‘Pass’ will appear on-
screen if the result is
within the tolerance
displayed on the
standard. If the result is
outside of the tolerance,
the instrument may need
to be re-calibrated.
Turbimeter
1.00 NTU
Next
Pass
01/01/2017 09:30:30
User 1
123456
Back
NTU
2.13
1 Check Standard Read
3
23
Follow the on-screen
prompts and insert the
Check Standards as
requested. Select the
Read button to perform
the measurement.
2
Summary of Contents for Turbimeter Plus
Page 7: ...4 Start Up Guide ...
Page 15: ...12 Read Menu ...
Page 22: ......
Page 23: ...20 Calibration Menu ...
Page 33: ...30 System Settings ...