Macro 900 Operation Manual
V3
© 2013 Palintest Ltd.
www.palintest.com
Page 140 of 147
24.
Appendix 4. Standard Electrodes Detailed Specification
Optical
Range
0
–
500.0% / 0
–
50.00 mg/L
Dissolved
Resolution
0.1% / 0.01mg/L
Oxygen
Accuracy
0 - 200%: ± 1% of reading. 200% - 500%: ± 10%
Range
0
–
200 mS/cm (0 - 200,000 µS/cm)
Conductivity
Resolution
3 Auto-range scales: 0
–
9999 µS/cm, 10.00
–
99.99 mS/cm, 100.0
–
200.0mS/cm
(EC)
Accuracy
± 1% of reading or ± 1µS/cm if greater
(see note 2)
Range
0
–
100,000 mg/L (ppm)
TDS*
Resolution
2 Auto-range scales: 0
–
9999mg/L, 10.00
–
100.00g/L
Accuracy
± 1% of reading or ± 1mg/L if greater
(see note 2)
Range
5
Ω•cm
–
1 M
Ω•cm
Resistivity*
Resolution
2 Auto-range scales: 5
–
9999
Ω•cm
, 10.0
–
1000.0 K
Ω•cm
Accuracy
± 1% of reading or ± 1
Ω•cm
if greater
(see note 2)
Range
0
–
70 PSU / 0
–
70.00 ppt (g/Kg)
Salinity*
Resolution
0.01 PSU / 0.01 ppt
Accuracy
± 1% of reading or ± 0.1 unit if greater
(see note 2)
Seawater
Range
0
–
50
σ
t
Specific
Resolution
0.1
σ
t
Gravity*
Accuracy
± 1.0
σ
t
Range
0
–
14 pH / ± 625mV
(see note 3)
pH
Resolution
0.01 pH / ± 0.1mV
Accuracy
± 0.1 pH / ± 5mV
Range
± 2000mV
(see note 3)
ORP
Resolution
0.1mV
Accuracy
± 5mV
Range
-5
˚
C
–
+50
˚
C (23
˚
F
–
122
˚
F)
Temperature
Resolution
0.1
˚
C/F
Accuracy
± 0.5
˚
C
*
Readings calculated from EC and temperature electrode values
Palintest Ltd reserves the right to change specifications without notice
Notes:
1.
The accuracy figures quoted throughout this document represent the
equipment’s capability at the calibration points at 25°C. These figures do not
take into account errors introduced by variations in the accuracy of
calibration solutions and errors beyond the control of the manufacturer that
may be introduced by environmental conditions in the field. Accuracy in the
field is also dependent upon
full calibration
and minimal time between
calibration and use.
2.
The EC electrode can be calibrated at various points for use in fresh,
brackish or salt water. The accuracy of the electrode, and therefore all
derived readings, relies upon the readings being within a reasonable range
of the calibration point. For taking measurements in fresh surface or ground
water, use Palintest MacroCal solution. If this is not available, use a third
party 1413µS/cm EC Calibration Standard. For taking readings in brackish or
salt water, use 12,880µS/cm (12.88mS/cm) EC Calibration Standard.
3.
The measurement of pH and ORP relies upon the ability of the electrode to