Macro 900 Operation Manual
V3
© 2013 Palintest Ltd.
www.palintest.com
Page 139 of 147
key whilst in this screen, any changes you have made will not be transferred to
the MAP 2000/2100.
Finally, refer to the relevant section of this manual and carry out a full two-
point (optical) or three-point (ISE) calibration of the new electrode.
YOUR
NEW ELECTRODE WILL NOT GIVE SENSIBLE READINGS UNTIL IT HAS
BEEN FULLY CALIBRATED.
Please note: changing an AUX Socket assignment will clear all the
calibration data for that socket.
If you subsequently remove an electrode, be sure to replace the blanking plug
and set the socket assignment back to EMPTY.