O
PERATION
M
ANUAL
VKL
10
/100
V
ERSION
V005-06-2020
27
9. Troubleshooting
Symptom
Presumable Cause
Elimination
dilution system "produces"
particles
nozzle, mixing chamber dirty
filter malfunction
inside channels dirty
outlet flow higher than total flow
environmental air is sucked in
cleaning
exchange filter
flush, or general
overhaul at plant
increase pre-pressure
dilution factor is not at nominal
value
over pressure or under-pressure at the
samplich oocation. Humidity- or
temperature impacts
inlet nozzle dirty
absence of isokinetic conditions
contact manufacturer
cleaning
take different outlet
nozzle, use correct
pressure
pressure gauge indicator not on
zero with VKL turned off
zero pont shifted
send in equipment