6
Verifying and calibrating measuring system
74
5200-en_V2.0_06/20
Overview of test points
Work through the following points, one after the other:
Item
Parameter
Detailed description
Check that aerosol tube and sensor unit
are unobstructed
-
Check that controller shows no leakage
Net leakage of controller
Maximum leakage permitted: 0.07 l/min
Check device for leakage
Net leakage of total system
Maximum leakage permitted: 0.49 l/min
Check aerosol sensor zero point
Cn
Correct value: 0.0
Set aerosol sensor offset
Offset
Permitted range of compensating volt-
age 2.0 - 3.0 V
Calibrate flow rate
Sensor flow rate
Maximum deviation permitted from ref-
erence measuring device value: 0.15 l/
min
Required only for initial commissioning
at a new location: Set average particle
velocity
Velocity calibrated
Calibrate particle size
Measured peak at
Maximum deviation permitted from set-
point raw channel: 0.5 channels
Calibrate particle velocity
Velocity at cursor
Maximum deviation permitted from
value indicated on Fidas calibration cer-
tificate: 0.5 m/s