3
Product description
5200-en_V2.0_06/20
19
3.4
Functional description
Analysis of particles in
ambient air
Fine dust measurement devices in the series
Fidas 200
are aerosol spec-
trometers certified in accordance with EN 16450. These devices continu-
ously analyze fine dust particles from the ambient air and simultaneously
calculate the immission values PM
2,5
and PM
10
, which require monitoring.
The values PM
1
, PM
4
, PM
tot
, the particle number concentration C
n
and the
particle size distribution are calculated and recorded. The number of parti-
cles is determined by means of optical light scattering.
Sampling and drying
The sampling head Sigma-2 draws in ambient air.
Volume flow rate: approx. 0.3 m
3
/h.
The aerosol passes through the sampling tube with drying section IADS and
into the aerosol sensor. The drying section dries the air that was sucked in,
thus preventing distortion of the measurement by any remaining moisture or
mist particles.
Weather data
The following weather-related factors have an impact on and control the dry-
ing effect:
• Air temperature
• Air pressure
• Humidity
The weather station measures this data. Additional data can be measured
and taken into consideration with some models:
WS300
WS500
WS600
Air temperature
Air temperature
Air temperature
Air pressure
Air pressure
Air pressure
Humidity
Humidity
Humidity
-
Wind direction
Wind direction
-
Wind speed
Wind speed
-
-
Precipitation
Scattered light analysis
The central element of the device is an optical aerosol sensor. Lorenz
‐
Mie
scattered light analysis of single particles is applied to determine the particle
size. The single particles move through an optically differentiated measure-
ment volume that is homogeneously illuminated with a polychromatic LED
light source. Each particle generates a scattered light impulse that is de-
tected at an angle of 85° to 95° degrees. The particle count is determined
based on the number of scattered light pulses. The level of the measured
scattered light pulse is a measure of the particle diameter.