6
Verifying and calibrating measuring system
5200-en_V2.0_06/20
73
Prerequisites
• Permitted temperature range during calibration: +5°C to +40°C.
• The device has to have been running for at least one hour before it can
be calibrated. It does not reach a thermally stable state until then.
• The device has to be in calibration mode:
Main menu > settings/calibration >
activate calib mode
activate calib modus
Screen
settings/calibration