6
Maintenance
6062-en_V2.4_02/21
59
Calibration dust MonoDust 1500
MonoDust
1500
A special dust with a very small monodisperse distribution and known
particle size is used to calibrate the device. Delivery of the device includes
one container of the calibration dust
MonoDust 1500
. Use only
MonoDust 1500
to calibrate the device.
When used properly, a single container is sufficient to calibrate the device
many times. Only a small amount of the dust is needed for the calibration
process.
MonoDust 1500
can be ordered from Palas or a sales partner.
The label on the container as well as the certificate for the calibration dust
indicate reference values (
setpoint raw channel
) needed to calibrate
the particle size.
This is how to use the calibration dust properly:
1.
Shake the closed container gently.
ð
Some of the particles swirl in the air inside the container and remain
in the air for a few minutes.
2.
Unscrew the container lid and hold the container horizontally next to
the sampling head.
Calibration arrangement: Container horizontal at sampling head.