7. Product Requirements and Major Parametric 

Description

Product Power Requirement

a) Power source: Rechargeable lithium battery3.7v/100mAh
b) Device safety classes: class II type BF

Product Main Technical Parametric description

a)Impulse frequency: 
TENS:20-100Hz

Problem

Analysis of cause

Recommended 
Solution

In order to avoid environmental pollution, please dispose of Aegis 
according to local environmental requirements. Do not discard in 
your standard waste disposal.

The device does 
not power up 
correctly

The buttons are 
not functioning 
correctly

The device does 
not emit any
"beep" sound

There is no 
output (electrical
pulses) or the 
output feels low

The battery is running 
low or running out

The button is damaged

The battery is running 
low or running out

The button is damaged

The speaker is 
damaged

The battery is running 
low or running out

The gel pad is 
damaged or dirty, has 
no stickiness or is 
making poor contact 
with the skin

Skin is too oily

Recharge the battery

Wipe the gel pad with 
moist fingers. Allow to 
dry before re-use. 
Alternatively, replace 
with a new gel pad

Wipe skin with a wet 
towel or alcohol wipe. 

Recharge the Battery

Contact the point of 
purchase

Recharge the Battery

Contact the point of 
purchase

Contact the point of 
purchase

11

10

The skin is 
strongly 
irritated/shows 
signs of 
redness

The treatment has 
been applied too long

The intensity is set too 
high

The gel pad is not 
making good contact 
with the skin

There may be an 
allergic reaction to the 
gel pad

The gel pad is dirty

Limit treatments to 15 
minutes

Decrease the intensity 
to a more comfortable 
level

Press the gel pad 
firmly against the skin 
to improve contact

Cease use of the 
device

Clean the gel pad

The device is damaged Allow to dry before 

using the device

Contact the point of 
purchase

Summary of Contents for KTR-2492

Page 1: ...uk Tel 0800 953 1666 E mail enquiries tower health co uk Transcutaneous Electronic Nerve Stimulator 20 12 09 L180 x W125mm 128G CE KTR 2492 24 4 p a in g o n e Aegis model KTR 2492 Aegis F a s t e f f...

Page 2: ...n obtained 3 Those who are or suspect they may be pregnant 4 Those with haemorrhage after acute trauma or fractures or those recovering from surgery 5 Those with skin particularly sensitive to heat or...

Page 3: ...atural pain relief mechanisms The device additionally delivers pleasant massage like stimulation Aegis can be worn discreetly under clothing and activated whenever needed via the dedicated remote cont...

Page 4: ...stic cover from side A of the gel pad Apply the gel pad to the Aegis unit as shown in the diagram Now remove side B from the gel pad Retain this plastic cover as it will be re used after treatment it...

Page 5: ...ly to cycle through different combinations of pulses to find a mode that you prefer There are 6 modes in Automatic function Each time you switch Function or mode the device intensity will reset to lev...

Page 6: ...h brand new gel pads Only use official Aegis gel pads with the Aegis device Only use one pair of pads per user of the device To purchase replacement gel pads contact Tower Health Ltd www tower health...

Page 7: ...ged The speaker is damaged The battery is running low or running out The gel pad is damaged or dirty has no stickiness or is making poor contact with the skin Skin is too oily Recharge the battery Wip...

Page 8: ...a 1060hPa c Transport environment requirements Environment temperature 25 55 Environment humidity 0 93 RH Atmospheric environment conditions 500hPa 1060hPa Indicates potential danger injury or serious...

Page 9: ...s such as electrocardiograph Do not use the product in strong direct sunlight or where there is high heat inflammable material electromagnetic radiation and or high humidity Do not disassemble repair...

Page 10: ...maging in operating rooms near active AF surgical equipment electrophysiology laboratories armored rooms or areas where short wave therapy equipment is used Do not use the system around strong electri...

Page 11: ...C 61000 4 4 3V 0 15 MHz 80 MHz 6 Vm in ISM bands between 0 15 MHz and 80 MHz 80 AM at 1KHz 3V 0 15 MHz 80 MHz 6 Vm in ISM bands between 0 15 MHz and 80 MHz 80 AM at 1KHz Conducted disturbances induced...

Page 12: ...217 Hz 2 2 0 3 0 3 0 2 0 3 0 3 28 28 9 1720 1845 1970 1700 1990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE 1 3 4 25 Band UMTS Pulse modulation 217 Hz 2 28 2450 2400 2570 Bluetooth WLAN 802 1 1 b g n RFID 2...

Reviews: