To return to Automatic function press and hold down the “F” button 
once again.

6. Timer

Each session lasts for 15 minutes. The remaining minutes of the 
session are displayed in the upper right corner of the screen on the 
remote control.  

The remote control will automatically turn off after the 15 minutes 
have elapsed. 

To begin another session simply re-activate the remote control, or 
press the “+” intensity button on the device.
The maximum recommended number of consecutive sessions is 4 
(60 minutes)

7. Powering off 

To end a session before 15 minutes have elapsed, simply press the 
power button      on either the power unit or the remote control.

8. Safety mode / Power cut-out

If the gel pads lose full contact with the skin the pulse intensity will 
automatically reduce to zero, as a safety feature. To re-commence 
treatment, re-position the gel pads firmly against the skin and press 
either the “+” or “–” buttons.

9. Storage

After use, remove Aegis from the body. Replace the plastic covers 
over the gel pads to protect them until the device is used again.

Replacing the gel pads

Gel pads can be refreshed by wiping them with moistened fingers. 
Allow to dry before using the device again. 

When the gel pads have lost their stickiness completely, they 
should be replaced with brand new gel pads.

Only use official Aegis gel pads with the Aegis device. 

Only use one pair of pads per user of the device. 

To purchase replacement gel pads contact Tower Health Ltd.
www.tower-health.co.uk 
Tel: 0800 953 1666
E-mail: [email protected]

Recharging the battery

The power unit and the remote control are both powered by lithium 
battery. These can be charged via the USB charging cable supplied.

The LED on the power unit will flash and a beep will sound to 
indicate when recharging is required.

On the remote control, a battery symbol will appear in the upper 
right corner of the screen to indicate that recharging is required.

Power unit charging time: Approximately 2 hours
Power unit battery life: Approximately 7 days, based on 15 minutes 
use per day.

Remote Control charging time: Approximately 2 hours
Power unit battery life: Approximately 20 days, based on 15 
minutes use per day.

Cleaning and maintenance

Keep the surface of gel pads clean. Avoid dust, oily matter and 
viscous matter. These can cause the adhesiveness of the gel pads 
to decline.

The control unit does not require maintenance or calibration. Do 
not attempt to open the control unit for any reason. If a fault is 
suspected, contact the point of purchase.

Storage

1) Keep Aegis out of reach of children
2) Do not store the device in places where it will be exposed to 
direct sunlight, high temperature or moisture.
3) Store the device in a dry and ventilated space

09

08

Summary of Contents for KTR-2492

Page 1: ...uk Tel 0800 953 1666 E mail enquiries tower health co uk Transcutaneous Electronic Nerve Stimulator 20 12 09 L180 x W125mm 128G CE KTR 2492 24 4 p a in g o n e Aegis model KTR 2492 Aegis F a s t e f f...

Page 2: ...n obtained 3 Those who are or suspect they may be pregnant 4 Those with haemorrhage after acute trauma or fractures or those recovering from surgery 5 Those with skin particularly sensitive to heat or...

Page 3: ...atural pain relief mechanisms The device additionally delivers pleasant massage like stimulation Aegis can be worn discreetly under clothing and activated whenever needed via the dedicated remote cont...

Page 4: ...stic cover from side A of the gel pad Apply the gel pad to the Aegis unit as shown in the diagram Now remove side B from the gel pad Retain this plastic cover as it will be re used after treatment it...

Page 5: ...ly to cycle through different combinations of pulses to find a mode that you prefer There are 6 modes in Automatic function Each time you switch Function or mode the device intensity will reset to lev...

Page 6: ...h brand new gel pads Only use official Aegis gel pads with the Aegis device Only use one pair of pads per user of the device To purchase replacement gel pads contact Tower Health Ltd www tower health...

Page 7: ...ged The speaker is damaged The battery is running low or running out The gel pad is damaged or dirty has no stickiness or is making poor contact with the skin Skin is too oily Recharge the battery Wip...

Page 8: ...a 1060hPa c Transport environment requirements Environment temperature 25 55 Environment humidity 0 93 RH Atmospheric environment conditions 500hPa 1060hPa Indicates potential danger injury or serious...

Page 9: ...s such as electrocardiograph Do not use the product in strong direct sunlight or where there is high heat inflammable material electromagnetic radiation and or high humidity Do not disassemble repair...

Page 10: ...maging in operating rooms near active AF surgical equipment electrophysiology laboratories armored rooms or areas where short wave therapy equipment is used Do not use the system around strong electri...

Page 11: ...C 61000 4 4 3V 0 15 MHz 80 MHz 6 Vm in ISM bands between 0 15 MHz and 80 MHz 80 AM at 1KHz 3V 0 15 MHz 80 MHz 6 Vm in ISM bands between 0 15 MHz and 80 MHz 80 AM at 1KHz Conducted disturbances induced...

Page 12: ...217 Hz 2 2 0 3 0 3 0 2 0 3 0 3 28 28 9 1720 1845 1970 1700 1990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE 1 3 4 25 Band UMTS Pulse modulation 217 Hz 2 28 2450 2400 2570 Bluetooth WLAN 802 1 1 b g n RFID 2...

Reviews: