Radiation Safety Guidelines

 

 

To ensure the health and safety of the operator and all others in the 
vicinity of operating x-ray inspection equipment, the following guidelines 
are recommended for establishing a basic radiation safety program. 

 

 

Note:  Federal, State, and some local government 
agencies may have more stringent regulations 
concerning the operation and use of equipment that 
produces ionizing radiation (x-rays).  The requirements 
of these governing agencies supersede the 
recommendations made by the manufacturer.

 

 

1.  A copy of the operating instructions should be kept at the 

machine at all times. 

 

2.  Personnel operating the equipment should be trained in the 

proper and safe operation of the machine. 

 

3.  Radiation surveys should be performed periodically to ensure 

that the amount of radiation being emitted by the machine is 
less than 0.5mR/hour.  We recommend performing a radiation 
survey: 

 

  Following initial installation 

– before placing the x-ray 

system into service. 

  Whenever the x-ray system is relocated. 

  Whenever the x-ray system receives a strong jolt (e.g., 

being dropped ore than one inch or struck sufficiently 
hard to dent or deform the exterior cabinet). 

  Whenever a leaded component (such as the x-ray 

generator, shielding, inspection tunnel curtains, LXDA or 
collimator cover) is removed for any reason or period of 
time. 

  Annually 

– from the date of the most recent radiation 

survey 

 

4.  All controls and indicators should be checked daily to ensure 

proper operation. 

 

5.  Do not operate the x-ray system if it is not in good repair.  Do 

not attempt to remove or bypass any controls (e.g. foot pedals 
and key-switches), interlocks, status indicators, leaded 
components (including the inspection tunnel curtains) or 
physical barriers.  In the event of failure or breakage, these 

Summary of Contents for XR 3700

Page 1: ...Ray Inspection System PN 8007 0542 120V PN 8007 0546 230V XR 3500 Real Time X Ray Inspection System PN 8007 0540 120V PN 8007 0544 230V This manual operation is for PACE s XR 3500 50kv unit and PACE...

Page 2: ...y Guidelines 6 XR 3700 Specifications 7 Features 7 System Operation 8 Set up 9 XR 3700 Initial Power up and Operational Test 10 XR 3700 Maintenance 11 XR 3700 Troubleshooting 12 Design Safety Standard...

Page 3: ...aware that the ionizing radiation x rays can constitute a distinct hazard if not employed in strict accordance with instructions provided in this manual for maximum operator safety Exposure to excessi...

Page 4: ...e a full understanding of how it works Additionally they should be familiar with established radiation safety exposure practices sanctioned by the National Bureau of Standards Handbook X ray Protectio...

Page 5: ...d constraints regarding the modification of x ray systems conforming to 21CFR mentioned above Typically an x ray system must be registered with the appropriate state agency by the physical possessor o...

Page 6: ...ount of radiation being emitted by the machine is less than 0 5mR hour We recommend performing a radiation survey Following initial installation before placing the x ray system into service Whenever t...

Page 7: ...PCB size 760mm x unlimited mm 30 x unlimited Focal spot 0 2mm 0 008 Focal spot to image plane distance 124mm 4 875 Adjustments Live or Capture video signal Frame averaging up to 32000 frames Video Ga...

Page 8: ...etermine the presence and degree of interlayer shifts At the onset of either conventional or small hole drilling inspection is used to qualify hole to pad alignment In addition to enhancing product qu...

Page 9: ...e bnc cable should be connected to the composite video input on the video monitor a bnc to RCA converter might be needed The power cord should be connected to a properly grounded 120 VAC power supply...

Page 10: ...as been returned to the off position The power light should blinks when the key switch is on and the cover lock is disengaged Then when the cover lock is engaged the power light turns steady 2 Make su...

Page 11: ...eo processor included with this system is preset to 8 frames averaging Averaging makes the x ray image less grainy and can be adjusted to improve image quality but with an increase in motion lag 8 Ima...

Page 12: ...t the system Power The power category covers all problems that include lack of power to a component or to the full system Power problems would include The system not powering up The x ray controller s...

Page 13: ...r Devices and Radiological Health Branch of the FDA as an Analytical X ray System As registered the system has a number of safety features provided to minimize any x ray scatter reaching the operator...

Page 14: ...ecific regulations regarding the monitoring of possible radiation leakage of industrial x ray cabinets are determined by the individual state or country There are a number of measures which can be tak...

Page 15: ...ption repair or replace any defective products at its facility or other location approved by it at no charge to user or provides parts without charge for installation by the user in the field at user...

Page 16: ...orporated to obtain the latest specifications The following are trademarks and or service marks of PACE Incorporated MD USA INSTACAL FUMEFLO HEATWISE PACEWORLDWIDE PERMAGROUND POWERMODULE and TEMPWISE...

Reviews: