Does the systems image processor boot up?

 

 

Does the X-ray monitor have a light gray screen or 
flicker?

 

 

X-Ray

:  The x-ray category covers the problems that affect the 

tube’s emission of x-rays.  The x-ray category includes: 

 

-  The red 

“X-Ray On” light is not turning on. 

 

-  There is no image on the monitor, but video signal is 
present. 
-  There are no x-rays being emitted from the x-ray source. 

 
 
 

 

Does an X-ray image appear on the monitor when 
the foot pedal is depressed?

 

Does the red 

“X-ray on” light turn on when the cover lock is engaged and 

the footswitch is depressed?

 

 

 
 

11. Model XR 3500/3700 Design Safety Standards and Safe Operating 

Practices

 

 
 

Because of the low operating anode power of the XR 3500/3700, real 
time x-ray inspection system, it has been registered with the Center for 
Devices and Radiological Health Branch of the FDA as an 

“Analytical X-

ray System”.  As registered, the system has a number of safety features 
provided to minimize any x-ray scatter reaching the operator. 

 

 

 

A.  Safety:

 

 

The X-Ray tube assembly is lead shielded with at least a ¾ inch space 
between collimator and image plane to minimize any x-ray scatter.  
The collimator insures that the x-ray beam size at the image plane is 
no more than a one-inch diameter circle.  The table assembly employs 
a lead acrylic view panel and additional lead shielding.  The highly 
sensitive x-ray camera permits the x-ray tube to operate at low power 
(anode voltage of 50/70 kV and anode current of 25 micro amps) 
resulting in minimal x-ray scatter.  All these safety features result in a 
structure with radiation scatter at 5 cm from any exposed surface less 
than 0.3 mill Roentgens per hour.  (Radiation exposure in an airplane 
at 30,000 feet during the day is greater than 0.4 mill Roentgens per 
hour.) 

Summary of Contents for XR 3700

Page 1: ...Ray Inspection System PN 8007 0542 120V PN 8007 0546 230V XR 3500 Real Time X Ray Inspection System PN 8007 0540 120V PN 8007 0544 230V This manual operation is for PACE s XR 3500 50kv unit and PACE...

Page 2: ...y Guidelines 6 XR 3700 Specifications 7 Features 7 System Operation 8 Set up 9 XR 3700 Initial Power up and Operational Test 10 XR 3700 Maintenance 11 XR 3700 Troubleshooting 12 Design Safety Standard...

Page 3: ...aware that the ionizing radiation x rays can constitute a distinct hazard if not employed in strict accordance with instructions provided in this manual for maximum operator safety Exposure to excessi...

Page 4: ...e a full understanding of how it works Additionally they should be familiar with established radiation safety exposure practices sanctioned by the National Bureau of Standards Handbook X ray Protectio...

Page 5: ...d constraints regarding the modification of x ray systems conforming to 21CFR mentioned above Typically an x ray system must be registered with the appropriate state agency by the physical possessor o...

Page 6: ...ount of radiation being emitted by the machine is less than 0 5mR hour We recommend performing a radiation survey Following initial installation before placing the x ray system into service Whenever t...

Page 7: ...PCB size 760mm x unlimited mm 30 x unlimited Focal spot 0 2mm 0 008 Focal spot to image plane distance 124mm 4 875 Adjustments Live or Capture video signal Frame averaging up to 32000 frames Video Ga...

Page 8: ...etermine the presence and degree of interlayer shifts At the onset of either conventional or small hole drilling inspection is used to qualify hole to pad alignment In addition to enhancing product qu...

Page 9: ...e bnc cable should be connected to the composite video input on the video monitor a bnc to RCA converter might be needed The power cord should be connected to a properly grounded 120 VAC power supply...

Page 10: ...as been returned to the off position The power light should blinks when the key switch is on and the cover lock is disengaged Then when the cover lock is engaged the power light turns steady 2 Make su...

Page 11: ...eo processor included with this system is preset to 8 frames averaging Averaging makes the x ray image less grainy and can be adjusted to improve image quality but with an increase in motion lag 8 Ima...

Page 12: ...t the system Power The power category covers all problems that include lack of power to a component or to the full system Power problems would include The system not powering up The x ray controller s...

Page 13: ...r Devices and Radiological Health Branch of the FDA as an Analytical X ray System As registered the system has a number of safety features provided to minimize any x ray scatter reaching the operator...

Page 14: ...ecific regulations regarding the monitoring of possible radiation leakage of industrial x ray cabinets are determined by the individual state or country There are a number of measures which can be tak...

Page 15: ...ption repair or replace any defective products at its facility or other location approved by it at no charge to user or provides parts without charge for installation by the user in the field at user...

Page 16: ...orporated to obtain the latest specifications The following are trademarks and or service marks of PACE Incorporated MD USA INSTACAL FUMEFLO HEATWISE PACEWORLDWIDE PERMAGROUND POWERMODULE and TEMPWISE...

Reviews: