Table 3: Guidance & Declaration - electromagnetic immunity concerning Conducted RF &
Radiated RF
Guidance & Declaration - Electromagnetic immunity
The
model
ProMate™ CL is intended for use in the electromagnetic environment specified below.
The customer or the user of the ProMate™ CL should assure that it is used in such an environment.
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic environment - guidance
Conducted
RF
IEC61000-4-6
Conducted
RF
IEC61000-4-6
Radiated
RF
IEC61000-4-3
3 Vrms
150 kHz to
80 MHz
6 Vrms ISM
frequency
band
3 V/m
80 MHz to
2.7 GHz
3V
6V
3V/m
Portable and mobile RF communications equipment
should be used no closer to any part of
the
ProMate
™
CL
, including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter. Recommended separation
distance
d
=1.2×P1/2 d=2×P1/2 d=1.2×P1/2 80 MHz to
800MHz
d
=2.3×P1/2 800 MHz to 2.7GHz where
P
is the
maximum output power rating of the transmitter
i
n
watts (W) according to the transmitter manufacturer and
d
i
s the recommended separation distance in meters
(m). Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey, (a)
should be less than the compliance level in each
frequency range. (b) Interference may occur
in
the
vicinity of equipment marked with the following symbol: