6.4.2 Preparation before cleaning Steps:
Tools: tray, soft brush, clean and dry soft cloth.
6.4.3 Cleaning
The cleaning should be performed no later than 24 hours after the operation. The cleaning can
be divided into automated cleaning and manual cleaning. Automated cleaning is preferred if
conditions permit.
6.4.3.1 Automated cleaning
•
The cleaner is proved to be valid by CE certification in accordance with EN ISO 15883.
•
There should be a flushing connector connected to the inner cavity of the product.
•
The cleaning procedure is suitable for the product, and the irrigating period is sufficient.
It is recommended to use a washer-disinfectant in accordance with EN ISO 15883. For the
specific procedure, please refer to the automated disinfection section in the next section
“Disinfection.”
1.
Remove the disposable prophy angle. Remove the outer sheath from the handpiece, and
place them onto a clean tray.
2.
Use a clean soft brush to carefully brush the head and back cover of the outer sheath until
the dirt on surface is not visible. Then use a soft cloth to dry the products and put them onto
a clean tray. The cleaning agent can be pure water, distilled water or deionized water.
3.
Disassembling steps
•
Remove the disposable prophy angle.
•
Remove the outer sheath from the handpiece.