Version 1.0 rev 14 Aug 2019
50
Balor
Table 12: Global Shutter Internal Triggering - Short Exposures Timing Parameters.
Parameter
Minimum
Maximum
Exposure
8 Rows
1 Frame + 20 Rows
Cycle Time (1/Frame Rate)
Ex 2 13 Rows
20,000 s
If the exposure time is greater than a frame read out time, the exposure starts first by pulsing the Global Clear. The
Reference frame is read out during the exposure such that the end of the Reference read out is coincident with the end
of the exposure. The Charge Transfer pulse then is applied to transfer charge from all pixels in the sensor. Finally, the
signal frame is then read as shown below in Figure 21.
Figure 21: Global Shutter Internal Triggering – Long Exposures.
Table 13: Global Shutter Internal Triggering - Long Exposures Timing Parameters.
Parameter
Minimum
Maximum
Exposure
1 Frame + 21 Rows
3600 s
Cycle Time (1/Frame Rate)
Ex 1 6 Rows
20,000 s