6.Technical Specifications
50
6.0000 A
0.2 + 0.05
10.000 A
[5]
0.250 + 0.05
True RMS
AC Current
[6]
60.000 mA, 600.00 mA,
6.0000 A, 10.000 A
[5]
20 Hz – 45 Hz
2 + 0.10
45 Hz – 2 kHz
0.50 + 0.10
2 kHz – 10 kHz 2.50 + 0.20
Resistance
[7]
600.00
Ω
1 mA
0.040 + 0.01
6.0000 k
Ω
1 mA
0.030 + 0.01
60.000 k
Ω
100
μ
A
0.030 + 0.01
600.00 k
Ω
10
μ
A
0.040 + 0.01
6.0000 M
Ω
1
μ
A
0.120 + 0.03
60.000 M
Ω
20
0 nA || 10 MΩ
0.90 + 0.03
100.00 M
Ω
200 nA || 10
MΩ
1.75 + 0.03
Diode Test
3.0000 V
[8]
1 mA
0.05 + 0.01
Continuity
1000
Ω
1 mA
0.05 + 0.01
Frequency
/Period
600 mV to 750 V
[9]
20 Hz – 2 kHz
0.01 + 0.003
2 kHz – 20 kHz 0.01 + 0.003
20 kHz – 200 kHz 0.01 + 0.003
200 kHz – 1 MHz 0.01 + 0.006
60 mA to 10 A
20 Hz – 2 kHz
0.01 + 0.003
2 kHz – 10 kHz 0.01 + 0.003
Capacitance
[10]
2.000 nF
200 nA
3 + 1.0
20.00 nF
200 nA
1 + 0.5
200.0 nF
2
μ
A
1 + 0.5
2.000
μ
F
10
μ
A
1 + 0.5
200
μ
F
100
μ
A
1 + 0.5
10000
μ
F
1 mA
2 + 0.5
Temperature
Temperature sensors under 2 categories supported -
thermocouple (ITS-90 conversion between B / E / J / K / N / R / S / T type), and
thermal resistance (RTD sensor conversion between PT100 and PT385 type)
[1]
Specifications are for 30-minute warm-up, "Low" measurement rate and calibration temperature
18
℃
- 28
℃
.
[2]
10% over range on all ranges, except 1,000 V DCV, 750 ACV, 10 A DCI, 10 A ACI,
100 MΩ resistance, and
10000 μF capacitance.
[3]
For each additional volt over ± 500 VDC add 0.02 mV of error.
[4]
Specifications are for amplitude of sine wave input > 0.5% of range. For inputs from 1% to 5% of range and
<50 kHz, add 0.1% of range extra error.
For 50 kHz to 100 kHz, add 0.13% of range extra error.
[5]
30 seconds OFF after 30 seconds ON is recommend for the continuous current that higher than DC 7 A or
AC RMS 7 A.
[6]
Specifications are for amplitude of sine wave input > 0.5% of range. 0.1% errors will be added when the
range of input sine wave is 1% to 5%.
[7]
Specifications are for 4–wire ohms function or 2–wire ohms using the relative operation of math. Without
relative operation, add ±0.20
Ω additional error in 2-wire ohms function.
[8]
Specifications are for the voltage measured at the input terminals. The 1 mA test current is typical. Variation
in the current source will create some variation in the voltage drop across a diode junction.