3.Functions and Operations
20
(4-wire Resistance)
2.
Connect the test lead.
I
2-wire Resistance
I
4-wire Resistance
3.
Set the range.
Press the Range softkey to set the range. Auto range automatically selects the range
for the measurement based on the input.
Note:
1000 V input protection is available in all ranges.
XDM3051:
20% over range for all ranges except 100 MΩ
range.
XDM3041: 1
0% over range for all ranges except 100 MΩ
range.
If the reading exceeds 105
MΩ
in 100 MΩ
range, "overload" will be displayed.
4.
Set the measurement speed.
Press the Speed softkey to switch between Low, Mid or High. See page 14,
Measurement Speed and Resolution
.
5.
Set the relative value.
(Advanced operation)
Press the Rel softkey to turn on or off the relative operation. For relative operation,
the multimeter subtracts the pre-specified value of REL operation from the actual