3.Functions and Operations
17
Note:
750 V input protection is available in all ranges.
XDM3051: 20% over range for all ranges except 750 V range.
XDM3041: 10% over range for all ranges except 750 V range.
If the reading exceeds 787.5 V in 750 V range, "overload" will be displayed.
4.
Set the measurement speed.
Press the Speed softkey to switch between Low, Mid or High. See page 14,
Measurement Speed and Resolution
.
5.
Set the relative value.
(Advanced operation)
Press the Rel softkey to turn on or off the relative operation. For relative operation,
the multimeter subtracts the pre-specified value of REL operation from the actual
measurement result and displays the result. See page 31,
Relative Value
.
Measuring DC Current
This section describes how to configure DC current measurements.
Operating Steps:
1.
Enable the DCI measurement.
Press
on the front panel to enter DCI measurement mode.
2.
Connect the test lead.
DC Current
3.
Set the range.
Press the Range softkey to set the range. Auto range automatically selects the range
for the measurement based on the input.
Note:
The multimeter uses two kinds of fuses for current protection: the 10 A current