4.Advanced User Guidebook
36
●Display
Push the
Utility
button, select
Function
in the right menu, select
Display
in the left
menu.
The description of
Display
Menu
is shown as the follows:
●
Adjust
Push the
Utility
button, select
Function
in the right menu, select
Adjust
in the left
menu.
The description of
Adjust
Menu
is shown as the follows:
Do Self Cal (Self-Calibration)
The self-calibration procedure can improve the accuracy of the oscilloscope under the
ambient temperature to the greatest extent. If the change of the ambient temperature is
up to or exceeds 5
℃
, the self-calibration procedure should be executed to obtain the
highest level of accuracy.
Before executing the self-calibration procedure, disconnect all probes or wires from the
input connector. Push the
Utility
button, select
Function
in the right menu, the function
menu will display at the left, select
Adjust
. If everything is ready, select
Self Cal
in the
Function Menu
Setting
Description
Function
Configure
Show the configure menu
KeyLock
Lock all keys. Unlock method: push
Trigger
Menu
button in trigger control area, then push
Force
button, repeat
3 times.
About
Show the version and serial number
Function Menu
Setting
Description
Function
Display
Show the display menu
BackLight
0% - 100%
Turn the
M
knob to adjust the backlight.
Graticule
Select the grid type
Menu Time
OFF, 5S – 30S
Turn the
M
knob to set the disappear time
of menu
Function Menu
Description
Self Cal
Carry out the self-calibration procedure.
Default
Call out the factory settings.
ProbeCh.
Check whether probe attenuation is good.