Page 44 of 55
USER MANUAL • Owandy-RX AC • 02/2020 • NXACEN160A
User Manual – Owandy-RX AC
DEVICE POWER SUPPLY
Type of power supply
single phase, alternate
Supply nominal voltage
115 V / 220V / 230V
Maximum voltage variation
±10%
Nominal current
3,5 A @ 220V / 230V
5,7 A @ 115V
Supply voltage frequency
50/60 Hz
Maximum line current (measured @ 70 kV, 8 mA, 3,2 s)
5,7 A @ 115V
Absorbed power
0,8 kVA at 230V
0,79 kVA at 220V
0,66 kVA at 115V
Apparent resistance
0,5 Ω at 220V / 230V
0,2 Ω at 115V
Protection fuses (F1 – F2 – F3 – F4)
F 8 A – 250 V at 220V / 230V
F 12,5 A – 250 V at 115V
Circuit protection fuses
(F5) – n° 1 630 mA – 125 V
(F6) – n° 1 500 mA – 125 V
ELECTRICAL CLASSIFICATION (IEC 60601-1)
Protection against electrical shock (insulation class)
Class I
Degree of protection against electrical Shock (applied part)
Type B (collimator cone)
Protection against harmful ingress of water or particulate matter
IP 20
Use with fl ammable anaesthetics
Not for use in presence of fl ammable anaesthetic
mixture with air, oxygen or nitrous oxide.
Sterilization and disinfection methods
The device is supplied not sterile and it must
not be subjected to sterilization
Operation mode
Continuous operation with intermittent X-ray
loading
MECHANICAL DATA
Total weight
19,5 kg (wall mounting)
50 kg (mobile version)
Weight of the tubehead
5,5 kg
Mechanical confi guration
Wall mounting, top and bottom / Mobile