Description of Parameters
Parameter
Setup Range or Item to be Selected
Units of Measure
Tag No.
Enter alphanumerics up to 8 characters.
Sensor Serial No.
Enter numerals up to characters.
Sensor type
Set the flowmeter type conforming to the flowmeter model code.
Enter alphanumerics up to 8 characters.
※
Date of manufacture
Numerals (the year represented by 2-digits)
Sensor material
Select one from SUS3, Hastelloy C, Monel, Tantalum,
Special, or Unknown.
Flange ratings
Select one from JIS 0K, JIS 20K, JIS 30K, ASME 50,
ASME 300, ASME 00, Special, or Unknown.
Description
Alphanumeric: accepts 5 characters max.
Generally used for entering brief comments.
Message
Alphanumeric: accepts 32 characters max.
Preamplifier display
Setup of the type of totalizer display in the flowmeter preamplifier.
Select one from
・
Total flow
・
Instantaneous flowrate
・
% instantaneous flowrate
・
% bar graph
Instantaneous flowrate
units
①
L/min
,
L/h
,
m
3
/min
,
m
3
/h
,
kL/min
,
kL/h
②
L/min
[
normal
],
L/h
[
normal
],
m
3
/min
[
normal
],
m
3
/h
[
normal
]
③
g/min
,
g/h
,
kg/min
,
kg/h
,
t/min
,
t/h
④
ton
(
US
)
/min
,
ton
(
US
)
/h
Select one from the available units having the same unit as
the attribute number in the "Totalized flow and factored pulse
units" setup.
Measurement units of
totalized flow and factored
pulse output
①
L
,
m
3
,
kL
②
L
[
normal
],
m
3
[
normal
]
③
g
,
kg
,
t
④
ton
(
US
)
Select one from the available units having the same unit as the
attribute number in the "Instantaneous flowrate units" setup.
Temperature units
One selected from
℃,°
F , or
K
Pressure units
Select one from Pa, kPa, MPa, mmHg, psi, bar, atm, or Torr.
(Each unit shows gauge pressure except for mmHg, atm and Torr.)
Meter factor
(value at 20
℃
)
Within
±
50% of nominal meter factor
L/P
Linear expansion
coefficient (
α
)
0
≦
"Linear expansion coeff (
α
)"
≦
0.00003
Default setting: 0.0000
Linear expansion
coefficient (
β
)
0
≦
"Linear expansion coeff (
β
)
≦
0.00003
Default setting: 0.0000
Cont'd
10. PARAMETER SETUP
Specifications of individual flowmeters are configured by establishing the following parameters.
Parameter are set through communications using a PC and OVAL Smart Communication Unit (Mode
EL230). (See the section under the topic "Wiring Connection." )
Since parameters are correctly established before the meter is delivered to the customer, no further
configuration is normally required.
For procedures to review and set up parameters, see the instruction manual for OVAL Smart
Communication Unit EL230.