![Oval Smart EX DELTA II Series Instructions Manual Download Page 19](http://html1.mh-extra.com/html/oval/smart-ex-delta-ii-series/smart-ex-delta-ii-series_instructions-manual_1662166019.webp)
D-582-3-E
Fig. .2
Fig. .
Fig. .3
Fig. .
Hex Socket Head
Screw
0
°
※
Latch
Cover
Shield Strip
Hex Soc. Head Screw
Sensor Leads
Turn
The preamplifier can be oriented to the desired
direction in 0
°
steps as shown in Fig. ..
NOTE: To change the orientation, hex key (JIS
B 8), nominal size , is required.
The preamplifier can be rotated by
loosening its four hex socked head
screws, but be sure to disconnect
sensor leads before attempting to
rotate it.
CAUTION:
Turning the preamplifier with
sensor leads in place may
damage the sensor.
Procedure
Refer to Section 5, "EXPLODED VIEW AND
PARTS LIST, FIXED TYPE" on pages and 2.
①
Turn off power.
②
Remove the terminal box cover (Fig. .2).
NOTE: If the preamplifier is of explosionproof,
disengage the latch (
※
) before
removing the cover.
③
Separate the shield strip (Fig. .3).
④
Disconnect sensor leads from the terminal
block.
⑤
Loosen a total of four hex socket head screws
securing the preamplifier's neck (Fig. .).
⑥
Turn the preamplifier to the desired direction,
exercising care not to force the sensor leads.
⑦
When the preamplifier orientation has been
set, assemble in the reverse order of removal.
6.10 How to Change Preamplifier Orientation
Summary of Contents for Smart EX DELTA II Series
Page 53: ...D 582 3 E 53 ...