E-922-1-E
11
5. FUNCTIONAL SETTINGS
Settings for the function of this instrument such as transmission rate, etc. are implemented by the hard
switches and soft switches. For the items of functional settings, refer to the Function Settings shown below.
5.1 Functional Settings
Function
Settings
Method for selection
Interface
RS422 (RS485)
—
Transmission rate
1200, 2400, 4800, 9600 (bps)
Hard switch
(DSW-1)
Data bit length
7, 8 (bit)
Soft switch
Delimiter setting
CR, CR+LF
Soft switch
Parity
None, Odd parity, Even parity
Soft switch
Stop bit
1, 2 (bit)
Soft switch
Local address
0 to 16 (stations)
Hard switch
(DSW-1)
Mode selection
Active mode, Polling mode
Soft switch
5.2 Hard Switch
The hard switch (hereinafter DSW-1) is located on the internal board. With DSW-1, you can set transmission
rate and local address.
DSW-1
LCD
Not used
Transmission
rate
0: 1200 bps
1: 2400 bps
2: 4800 bps
3: 9600 bps
Local address
0: Station 1
1: Station 2
F: Station 16
1
2
0
2
2
1
3
2
0
4
2
1
5
2
0
7
2
2
8
2
3
6
2
1
(SW number)
Internal board
• DSW-1 Setting
to
DSW-1
LCD
Not used
Transmission
rate
0: 1200 bps
1: 2400 bps
2: 4800 bps
3: 9600 bps
Local address
0: Station 1
1: Station 2
F: Station 16
1
2
0
2
2
1
3
2
0
4
2
1
5
2
0
7
2
2
8
2
3
6
2
1
(SW number)
Internal board
• DSW-1 Setting
to
CAUTION: Set DSW-1 appropriately to meet
your operating environment.
If you set it incorrectly,
malfunction or damage of
instrument may occur.
Fig. 5.2
Fig. 5.1