E-822-9-E

4

4.INSTALLATION

4.1 Installation Location

Select an installation location where

1

Mechanical vibration, shock and corrosive 

gases least exist.

2

Air is dry and temperature at room 

temperature and stable.

NOTE: Although  the  manufacturer  guar-

antees  stated  performance  at 
ambient temperatures up to 50

, it 

is recommended that the instrument 
be placed in service at room temper-
ature.

3

Provide a sufficient working space behind 

the instrument - at least 50 centimeters from 
the back of the instrument - to facilitate wiring 
and maintenance.

4.2 Mounting Panel

1

Use  a  rigid  steel  sheet  with  a  minimum 

thickness  of  1.6  millimeters.  Suggested 
thickness is approximately 3.2 mm.

2

If it is desired to mount instruments side by 

side, dimensions in Fig.4.1 are suggested.

3

Recommended mounting height ranges are 

given in Fig.4.2.

4.3 Installation Procedure

1

Front mount the steam flow computer through 

the opening cut in the panel.

2

Pass the enclosure mounting fittings furnished 

through  the  holes  at  the  bottom  of  the 
enclosure and, con-firming that the instrument 
is positioned on a level plane, secure the 
instrument to the panel with the mounting 
fittings (see Fig. 2).

Fig. 4.1 Panel Cutout

Fig. 4.2 Mounting Height

95min.

68

−0.7

   0

138

+1  0

20

0

min.

OPTIMUM RANGE

ALLOWABLE RANGE

600

1800

450

2200

95min.

68

−0.7

   0

138

+1  0

20

0

min.

OPTIMUM RANGE

ALLOWABLE RANGE

600

1800

450

2200

Unit

mm

Unit

mm

Summary of Contents for EL4061

Page 1: ...e throughout its life this manual offers the operator the necessary installation operation and maintenance information Be well familiar with these instructions before you place the computer in service...

Page 2: ...TION 8 10 1 Preparation Before Operation 8 10 2 Operation 8 11 TROUBLESHOOTING 8 12 OPERATION FLOW CHART 9 13 PRODUCT CODE EXPLANATION 11 14 TOTAL COUNTER AND OUTPUT UNITS 11 15 GENERAL SPECIFICATIONS...

Page 3: ...S 3 OUTLINE DIMENSIONS OPERATION INDICATOR LED RED 7 DIGIT TOTAL COUNTER UNITS OF MEASURE NAMEPLATE PRESSURE SETTING SWITCH UNITS OF MEASURE MPa SEFT TEST BUTTON TAG NO NAMEPLATE ENCLOSURE MTG FITTING...

Page 4: ...a minimum thickness of 1 6 millimeters Suggested thickness is approximately 3 2 mm 2 If it is desired to mount instruments side by side dimensions in Fig 4 1 are suggested 3 Recommended mounting heig...

Page 5: ...ve interference 3 Using crimp style terminals for wiring connections ensure good electrical connections Terminals are found on the back of the computer 4 Good practice is to provide an allowance of 30...

Page 6: ...TTING RANGE 0 00 to 1 99MPa PRESSURE SETTING SWITCH 7 DIGIT TOTAL COUNTER LED TEST CHECK RUN SELECT SWITCH FRONT PANEL SIZE NOM DIA SETTING SWITCH ONE SHOT MULT CIRCUIT SINGLE CHIP MICROPROCESSOR 6 OV...

Page 7: ...7567 4 C 150 2 422 5 40 0 04363 D 200 7 021 6 40 0 04556 E 250 13 54 7 50 0 09655 F 300 23 24 NOTE Separate rotary switch positions are allocated because positions 2 and 3 5 and 6 7 and 8 have differe...

Page 8: ...CK mode CHECK mode operation is described on Sec 12 10 2 Operation 1 Turning on power The LED will light up about a second later and stays on for about two seconds before the RUN mode is indicated 2 A...

Page 9: ...HECK MODE FLOW CHART CHECK MODE LED blinks for 2 sec Total counter counts 3 pulses LED lights in synchronism LED blinks for 2 sec LED lights for pre set number at in tervals of 0 5 sec LED lights for...

Page 10: ...rotary switch position currently set up Postitions A B C D E and F are indicated by blinking 10 11 12 13 14 and 15 times respectively NO NO YES YES Meter error set or not Shows whether positive correc...

Page 11: ...rotary switch establishes value which agrees with meter size For rotary switch positions see Section 8 2 Setting of Meter Size nominal bore Meter Error Internal jumper establishes corrections in a ran...

Page 12: ...ove Always N N Always N Output item 2 Pulse output N N None Always N N Always N Output item 3 Analog output N N None A 2 1 to 5V A 5 4 to 20mA A P 0 to 10mV A Q 10 to 50mV A R 0 to 100 A Z Z Special A...

Page 13: ...erter output provided 0 to 10mV 4 F I converter output provided 4 to 20mA 5 F I converter output provided 10 to 50mA 7 F I converter output provided 0 to 100 A 9 Other than above 1 Always 1 Instrument...

Page 14: ...E 822 9 E All specifications are subject to change without notice for improvement 2019 04 Revised E 822 9 E 1...

Reviews: