CAUTION
Lack of instruction
Injuries, malpositions, illness of the user due to incorrect information
►
The qualified personnel is responsible for making sure that the user and/or attendant/nursing staff has under
stood the proper adjustment, use, maintenance and care of the belt system.
►
In particular, ensure that the user and/or attendant/care staff knows how to quickly loosen and open the
product to avoid delays in case of emergency.
Information on the settings can be found in the manufacturer’s instructions for use accompanying each product.
6.11 Adjusting the tray
The tray slides onto the arm supports.
If the tray holder was not adapted to the arm support height, this can be done with the help of the enclosed instruc
tions for use or assembly instructions.
7 Delivery
7.1 Final inspection
A final check must be carried out before the wheelchair can be handed over:
•
Are all options installed according to the order form?
•
Are the drive wheels correctly positioned?
•
Do the quick-release axles rotate freely and lock securely?
•
Have the tyres been correctly inflated?
INFORMATION: The correct tyre pressure is printed on the tyre sidewall. For drive wheels with high-
pressure tyres, the minimum air pressure is 7 bar.
•
Only after adjustment: Have the brakes (wheel locks) been adjusted correctly?
•
Only after adjustments: Has the respective caster journal angle been adjusted vertically?
•
Only after adjustments: Has the anti-tipper been adjusted correctly?
7.2 Transport to the customer
NOTICE
Use of unsuitable packaging
Damage to the product caused by transportation using incorrect packaging
►
Use only the original packaging for delivery of the product.
The wheelchair should be transported to the user in disassembled state using the outer packaging.
7.3 Handing over the product
WARNING
Lack of instruction
Tipping over, falling of the user due to lack of knowledge
►
Instruct the user or the attendant in the proper use of the product when handing it over.
The following steps must be performed for the safe delivery of the product:
•
Conduct a sitting test with the user of the product. Pay special attention to proper positioning according to
medical considerations.
•
The user and any attendants must be instructed in the safe use of the product. In particular, the enclosed
instructions for use (user) are to be used.
•
The instructions for use (user) must be issued to the user/attendant during handover of the wheelchair.
•
Depending on equipment:
The supplied instructions for use for accessories have to be handed over in addi
tion.
8 Maintenance and repair
The manufacturer recommends regular maintenance of the product every
12 months
.
Further information on caring for the product and on maintenance/repair can be found in the instructions for use
(user).